Hamilton C1

Alarm: 231006 Oxygen supply failed

On this page

Asset Type

Ventilator

Manufacturer

Hamilton

Model

C1

What This Guide Helps With

This guide helps troubleshoot Alarm 231006 Oxygen supply failed on the Hamilton C1 ventilator. This alarm may occur when oxygen supply pressure or O2 control flow is too low, which can affect the ventilator’s ability to deliver the expected FiO2.

The issue may be caused by a failed wall oxygen source, closed or low cylinder, damaged O2 hose, loose inlet connection, restricted inlet filter, transport setup problem, or an internal oxygen control or flow sensing fault.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm the Oxygen Source Is Available

Check the O2 Hose

Check the Oxygen Inlet Connection

Check Oxygen Supply Pressure

Inspect the O2 Inlet Filter

Rule Out Transport or Setup Issues

Test With a Known-Good Oxygen Source and Hose

Evaluate for Internal O2 Control or Flow Sensing Fault

If the Problem Persists

If the oxygen wall source, cylinder or regulator, O2 hose, inlet connection, supply pressure, and inlet filter have been checked and the alarm remains, external and common causes have been ruled out.

At that point, the Hamilton C1 should be removed from service, labeled Out of Service, and sent for repair, bench evaluation, or manufacturer service support.

Persistent oxygen supply failed alarms may indicate a device-related oxygen control, inlet regulation, valve, pneumatic, or flow sensing issue.

Clinical Use Tip

Never troubleshoot this alarm on an active patient while oxygen delivery is uncertain. Move the patient to a backup ventilator or approved backup ventilation method first, then continue equipment troubleshooting only after patient safety is controlled.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Hamilton C1 ventilator reported Alarm 231006 Oxygen supply failed during use. Staff reported concern that delivered FiO2 may not match the set oxygen concentration."

Cause

What was observed during troubleshooting.

Example:
"Oxygen supply path issue suspected. Wall oxygen source, O2 hose, inlet connection, supply pressure, O2 inlet filter, and oxygen source availability were checked. Alarm persisted with known-good external oxygen supply, indicating likely internal O2 control or flow sensing fault."

Resolution

What action was taken.

Example:
"Removed ventilator from patient use, labeled Out of Service, and sent for bench evaluation/repair. Documented alarm behavior and external oxygen supply checks performed."

Helpful Details to Include (If Known)

Final Thought

Alarm 231006 Oxygen supply failed should be handled as a patient safety issue first and a troubleshooting issue second. Start with the oxygen source, hose, inlet connection, pressure, and filter before suspecting the ventilator. If verified external oxygen supply does not clear the alarm, remove the Hamilton C1 from service and escalate for repair.

Clear CCR documentation helps show what was checked, what was ruled out, and why repair escalation was appropriate.

That is successful troubleshooting.