Hamilton C1

Alarm: 244001 Loss of external power

On this page

Asset Type

Ventilator

Manufacturer

Hamilton

Model

C1

What This Guide Helps With

This guide helps troubleshoot Alarm: 244001 Loss of external power on the Hamilton C1 ventilator. This alarm indicates the ventilator has lost external AC power and may be running on battery power. If external power is not restored, battery runtime may become limited and patient ventilation could be affected.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check the AC Power Cord

Check the Wall Outlet

Check the External Power Supply or Adapter

Check for Movement-Related Power Loss

Check Battery Status

Check Facility Power Conditions

Confirm Alarm Clears and External Power Is Stable

If the Problem Persists

If Alarm: 244001 Loss of external power continues after checking the power cord, wall outlet, external power supply or adapter, facility power, and battery status, external and common causes have been ruled out.

Persistent loss of external power may indicate an internal power input, charging, battery interface, or power management failure.

Remove the ventilator from service.

Label it Out of Service.

Send it for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot a power alarm for an extended period while the ventilator is supporting an active patient. Restore external power immediately, monitor battery status closely, and move the patient to a backup ventilator or approved alternate ventilation method before continuing troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Hamilton C1 ventilator reported Alarm: 244001 Loss of external power. Staff stated ventilator was running on battery while connected to wall power."

Cause

What was observed during troubleshooting.

Example:
"AC power cord was loose at the external power connection. Wall outlet tested good, and battery status was acceptable after external power was restored."

Resolution

What action was taken.

Example:
"Reseated AC power cord, verified external power indicator returned, confirmed battery charging, and monitored ventilator with no recurrence of alarm. Device returned to service."

Helpful Details to Include (If Known)

Final Thought

Alarm: 244001 Loss of external power should be treated as a patient safety issue because the ventilator may only be supported by battery power. Start with the simple external causes first: power cord, outlet, adapter, facility power, and battery status. If the alarm persists after those checks, remove the ventilator from service and escalate for repair. Clear CCR documentation shows what was tested, what was ruled out, and why the final decision was made.

That is successful troubleshooting.