Hamilton C1

Alarm: 244013 Battery 2: Defective

On this page

Asset Type

Ventilator

Manufacturer

Hamilton

Model

C1

What This Guide Helps With

This guide helps troubleshoot Alarm: 244013 Battery 2: Defective on a Hamilton C1 ventilator. This alarm indicates Battery 2 has failed, is not usable for backup power, or cannot be trusted for safe operation during power loss, transport, or accidental disconnection from AC power.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm External Power Is Stable

Check Battery 2 Seating

Inspect Battery 2 and Battery Contacts

Replace Battery 2 With an Approved Compatible Battery

Confirm Battery 2 Is Recognized and Charging

Compare Battery Slot Behavior if Needed

Check for Related Power or Battery Alarms

If the Problem Persists

If Alarm: 244013 Battery 2: Defective remains after confirming stable AC power, reseating Battery 2, inspecting contacts, and installing an approved known-good replacement battery, the external and common battery causes have been ruled out.

The remaining issue may involve the ventilator-side Battery 2 bay, battery contacts, charging circuit, battery manager, or internal power management hardware.

Remove the device from service.

Label it Out of Service.

Send it for repair or bench evaluation.

Do not rely on the ventilator for patient use until the Battery 2 fault is corrected and backup power operation is verified.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device first. A defective backup battery can become critical during transport, outlet failure, or accidental power disconnection.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Hamilton C1 ventilator displayed Alarm: 244013 Battery 2: Defective. Staff reported Battery 2 warning was present during startup and battery status showed Battery 2 was not usable."

Cause

What was observed during troubleshooting.

Example:
"Battery 2 had a permanent failure. Battery contacts were inspected and no external damage was found. Alarm cleared after Battery 2 was replaced with an approved compatible battery."

Resolution

What action was taken.

Example:
"Replaced Battery 2, confirmed replacement battery was recognized and charging normally, verified no active battery alarms remained, and completed operational check. Ventilator returned to service."

Helpful Details to Include (If Known)

Final Thought

Battery alarms on ventilators should always be treated as patient safety concerns. Start with safe power, confirm the exact alarm, inspect the battery and contacts, replace the failed battery, and verify the replacement is recognized and charging. If the alarm remains with a known-good battery, escalate the ventilator for repair instead of continuing bedside troubleshooting.