Hamilton C6

Air or O₂ Gas Supply Failure

On this page

Asset Type

Ventilator

Manufacturer

Hamilton

Model

C6

What This Guide Helps With

Troubleshooting oxygen-supply alarms, incorrect oxygen delivery, restricted ambient-air intake, damaged hoses, source-pressure problems, or turbine-related airflow failure.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot a suspected gas-supply failure while the HAMILTON-C6 is actively supporting a patient.

Expected outcome: The patient is safely supported without depending on a ventilator with an uncertain gas supply.

Continue Clinical Engineering troubleshooting only after the affected ventilator has been removed from patient use.

2. Confirm the Exact Alarm or Failure

Record the complete alarm wording and determine whether the reported problem involves:

The HAMILTON-C6 uses a turbine to draw ambient air and uses an external high-pressure oxygen source. Therefore, a reported “air supply failure” may involve the ambient-air inlet or turbine rather than a disconnected wall-air hose.

Expected outcome: The failure is identified as an oxygen-source issue, ambient-air restriction, setup problem, or possible internal airflow failure.

3. Inspect the Ventilator Externally

Check the ventilator for:

Move the ventilator to a clean, open test location with adequate clearance around its air openings.

Expected outcome: The ventilator has unobstructed airflow and no obvious physical damage.

If clearing an external obstruction resolves the issue and the ventilator passes all required testing, stop troubleshooting.

4. Verify the Oxygen Source

Confirm that the oxygen source is available and appropriate:

Do not use an unverified source merely because gas can be heard flowing.

Expected outcome: A stable, verified high-pressure oxygen source is available to the ventilator.

If the source is defective, remove that outlet, regulator, or cylinder from use and notify the appropriate department.

5. Inspect the Oxygen Hose and Fittings

Disconnect the oxygen hose from service pressure before inspecting it.

Check for:

Reconnect both ends securely. Substitute a known-good, compatible oxygen hose when possible.

Expected outcome: Oxygen reaches the ventilator through a correctly connected, undamaged hose.

If replacing the hose resolves the alarm, complete operational testing and stop troubleshooting.

6. Confirm the Correct Gas Connection

Verify that the hose is connected to the ventilator’s oxygen inlet and to a verified oxygen source.

Check that:

The manufacturer identifies high-pressure oxygen from a central supply or cylinder as the C6 oxygen source.

Expected outcome: The correct gas is connected to the correct inlet with secure, gas-specific fittings.

7. Check the Ambient-Air Intake Area

Because the C6 uses an internal turbine for airflow, inspect the accessible ambient-air intake and external filter area without opening the ventilator.

Check for:

Replace only user-accessible filter components when permitted by facility procedure and manufacturer instructions.

Expected outcome: Ambient air can enter the ventilator without external restriction.

Do not open the ventilator or attempt turbine disassembly during initial troubleshooting.

8. Restart the Ventilator Off the Patient

With the ventilator connected to verified AC power and a verified oxygen source:

Record any displayed alarm number, technical code, or message exactly as shown.

Expected outcome: The ventilator completes startup without a gas-supply or turbine-related fault.

If the alarm returns, continue troubleshooting.

9. Review Oxygen and Ventilation Settings

Using a test lung, review:

Determine whether the failure occurs:

Do not change clinical settings merely to suppress an alarm. Use controlled test settings appropriate for bench evaluation.

Expected outcome: The issue is reproduced under controlled conditions and is not caused by an unrealistic or unintended setup.

10. Perform the Required Preoperational Check

Connect the ventilator to a manufacturer-compatible test circuit and test lung. Run the applicable preoperational checks according to the facility’s approved procedure.

Verify:

Hamilton provides C6-specific preoperational-check resources as part of its official training materials.

Expected outcome: The ventilator passes its preoperational checks without gas-supply or airflow faults.

Do not return the device to service if any required test fails.

11. Verify Delivered Oxygen and Ventilator Performance

Using calibrated test equipment, compare the set oxygen concentration with the measured delivered concentration at several representative settings.

Also verify:

Follow the facility’s performance limits and current manufacturer service documentation.

Expected outcome: Delivered oxygen and ventilation performance remain within approved limits.

If oxygen concentration is inaccurate despite a verified source, hose, fittings, and setup, stop troubleshooting and escalate the device.

12. Compare with Known-Good External Components

When available, repeat testing using:

Change one component at a time so the cause can be identified.

Expected outcome: The failure either follows an external component or remains with the ventilator.

If the fault remains with the ventilator after all external components are verified, internal evaluation is required.

If the Problem Persists

If the HAMILTON-C6 continues to display an oxygen-supply, turbine, airflow, or delivered-FiO₂ fault after the source, hose, fittings, ambient-air intake, settings, circuit, and test equipment have been verified, common external causes have been ruled out.

The issue may involve an internal oxygen-control component, pressure sensor, turbine assembly, internal pneumatic pathway, control electronics, or calibration fault.

The ventilator should be:

Do not open the pneumatic system or replace internal components unless trained, authorized, and equipped to complete the required calibration and performance verification.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot uncertain gas delivery on an active patient. Transfer the patient first, maintain independent oxygenation and ventilation monitoring, and test the ventilator only with a test lung in a controlled environment.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported that the HAMILTON-C6 displayed an oxygen-supply alarm and could not maintain the selected FiO₂."

Cause

What was observed during troubleshooting.

Example:
"Testing found a damaged oxygen hose fitting that caused an unstable high-pressure oxygen connection."

Resolution

What action was taken.

Example:
"Replaced the oxygen hose, connected the ventilator to a verified oxygen source, completed preoperational and performance testing, and returned the unit to service."

Helpful Details to Include (If Known)

Final Thought

Gas-supply troubleshooting must begin with patient safety and verification of the external oxygen source, hose, fittings, ambient-air intake, and setup. When the fault remains after controlled testing, remove the ventilator from service and escalate it rather than assuming it is safe to use. Clear CCR documentation preserves the reasoning, findings, and final equipment status.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide