Hamilton C6

CO₂ or SpO₂ Monitoring Module Not Working

On this page

Asset Type

Ventilator

Manufacturer

Hamilton

Model

C6

What This Guide Helps With

Troubleshooting missing CO₂ or SpO₂ data caused by disabled monitoring, loose connections, damaged sensors, contaminated adapters, poor signal, or configuration issues.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot unreliable CO₂ or SpO₂ monitoring while the ventilator is supporting a patient when either parameter is clinically required.

Expected outcome: The patient is safely supported without relying on the affected monitoring function.

Continue Clinical Engineering troubleshooting only when the device can be evaluated safely.

2. Confirm Which Monitoring Function Failed

Determine whether the problem affects:

Record the exact alarm or message shown on the ventilator.

Expected outcome: The failure is clearly identified before components are disconnected or replaced.

3. Verify the Monitoring Option Is Installed

Confirm that the ventilator is configured with the applicable CO₂ or SpO₂ monitoring hardware and licensed option.

Do not assume every HAMILTON-C6 has both monitoring capabilities installed. Hamilton identifies CO₂ and SpO₂ as optional integrated accessories on applicable configurations.

Expected outcome: The requested monitoring function is supported by the ventilator’s installed configuration.

If the option is not installed, stop troubleshooting and report the configuration limitation.

4. Confirm the Sensor Is Enabled

From the ventilator display, open:

System > Sensors > On/Off

Verify that the appropriate selection is enabled:

The C6 requires CO₂ and SpO₂ monitoring to be individually enabled before their data becomes available.

Expected outcome: The required sensor is enabled and its monitoring fields become available.

If enabling the sensor resolves the issue, verify stable operation and stop.

5. Inspect the External Connection

Trace the affected sensor from the patient end back to the ventilator.

Check for:

Disconnect and reconnect the sensor using proper connector alignment. Do not twist or force the connector.

Expected outcome: The sensor and adapter are fully seated with no visible damage.

If reseating restores monitoring, perform a functional verification and stop.

6. Power-Cycle the Monitoring Accessory

When safe and permitted by facility procedure:

For CO₂, allow any displayed warm-up period to complete before judging performance.

Expected outcome: The ventilator recognizes the accessory and begins displaying valid monitoring information.

7. Check the CO₂ Sensor and Airway Adapter

For a CO₂ complaint:

The C6 supports external CO₂ sensor and airway-adapter assemblies, including mainstream configurations.

Expected outcome: The adapter is clean, correctly assembled, and capable of passing a stable optical CO₂ signal.

If replacing the adapter restores the capnogram and values, verify operation and stop.

8. Check Sidestream Components, If Equipped

For a sidestream CO₂ configuration:

Do not attempt to clean or reuse single-use sampling components.

Expected outcome: The sampling path is open, dry, and properly connected.

9. Check the SpO₂ Adapter and Patient Sensor

For an SpO₂ complaint:

Hamilton C6 configurations may use an external SpO₂ adapter containing the pulse-oximetry hardware, with a compatible patient sensor connected to it.

Expected outcome: All SpO₂ components are compatible, undamaged, and fully connected.

10. Rule Out a Poor SpO₂ Signal

When performing an operational check:

A numerical SpO₂ value without a stable plethysmogram may indicate poor signal quality rather than ventilator failure.

Expected outcome: A stable plethysmogram, pulse rate, and SpO₂ value appear.

11. Substitute Known-Good External Components

Use compatible, verified components to isolate the failure.

For CO₂, substitute one component at a time:

For SpO₂, substitute one component at a time:

Do not substitute incompatible brands or technologies merely because the connectors appear similar.

Expected outcome: The failed external component is identified without replacing multiple parts unnecessarily.

If a known-good accessory resolves the issue, replace the defective component and stop.

12. Review Alarms and Perform Available Checks

Review the alarm history and displayed messages for indications such as:

Perform only the user-accessible calibration, zeroing, or test functions authorized by the operator documentation and facility procedure. Do not perform internal adjustments.

Expected outcome: Required external calibration or sensor checks complete successfully.

13. Restart the Ventilator Off Patient

After the patient has been safely transferred and configuration information has been recorded:

Expected outcome: The ventilator recognizes the monitoring accessory after a controlled restart.

If the accessory still is not recognized, proceed to escalation.

If the Problem Persists

If compatible known-good sensors, adapters, cables, sampling components, settings, and external connections have been verified, common external causes have been ruled out.

The problem may involve:

The ventilator should be:

Do not open the ventilator or replace internal assemblies during routine floor troubleshooting. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Do not depend on missing or intermittent integrated CO₂ or SpO₂ values during patient care. Provide independent monitoring or move the patient to another verified ventilator before disconnecting sensors, restarting the C6, or performing tests that could affect therapy.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported that the HAMILTON-C6 displayed no CO₂ waveform or SpO₂ value after the monitoring accessory was connected."

Cause

What was observed during troubleshooting.

Example:
"Inspection found a damaged SpO₂ adapter cable that failed with movement, while the ventilator operated normally with a compatible known-good adapter and sensor."

Resolution

What action was taken.

Example:
"Replaced the defective SpO₂ adapter, enabled SpO₂ monitoring, verified a stable plethysmogram and value, and returned the ventilator to service."

Helpful Details to Include (If Known)

Final Thought

Patient monitoring problems should be approached by protecting the patient first, identifying the exact failed function, and checking configuration, connections, accessories, and signal quality before suspecting internal failure. Clear CCR documentation supports safe repair decisions and prevents repeated troubleshooting.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide