Hamilton C6

Flow Sensor Calibration Failure

On this page

Asset Type

Ventilator

Manufacturer

Hamilton

Model

C6

What This Guide Helps With

Troubleshooting failed flow-sensor calibration caused by incorrect sensor selection, loose tubing, circuit leaks, moisture, improper seating, or defective external components.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform flow-sensor calibration while the ventilator is connected to a patient.

Expected outcome: The patient is safely supported without relying on inaccurate flow, volume, or pressure measurements.

The C6 identifies missing or unusable flow-sensor calibration data as a high-priority condition.

2. Confirm the Reported Failure

Place the ventilator in Standby and open:

System > Tests & calib > Flow sensor

Determine whether:

Record the exact displayed message and whether the failure occurs every time.

Expected outcome: The failure condition is clearly identified before components are changed.

3. Verify the Selected Patient Group

Confirm that the installed flow sensor matches the selected patient group:

A mismatch between the selected patient group and installed flow sensor will cause calibration to fail.

Expected outcome: The correct sensor type and patient category are selected.

If correcting the patient group resolves calibration, complete the remaining preoperational checks and stop.

4. Inspect the Flow Sensor

Remove the sensor from clinical use and inspect it externally for:

Do not insert tools or compressed air into the sensor.

When active humidification is used, accumulated water can interfere with proximal flow measurement. Hamilton recommends positioning the flow sensor at least 45 degrees relative to the floor to reduce water accumulation.

Expected outcome: The sensor is clean, dry, undamaged, and appropriate for the circuit.

5. Check the Flow-Sensor Tubing and Connections

Inspect both flow-sensor pressure tubes from the sensor to the ventilator.

Confirm that:

Reseat each connection firmly without forcing it.

Expected outcome: Both differential-pressure pathways are open and securely connected.

If reseating the tubing resolves calibration, complete the preoperational check and stop.

6. Inspect the Breathing Circuit for Large Leaks

Check the complete test circuit between the ventilator and flow sensor, including:

Reconnect any loose component and replace visibly damaged tubing or accessories.

Hamilton identifies circuit disconnections and large leaks, including leaks around the humidifier, as causes of failed flow-sensor calibration.

Expected outcome: The circuit is complete, correctly assembled, and free of obvious leaks.

7. Check the Expiratory Valve Set

Remove and reinstall the expiratory valve set according to the approved user-level setup procedure.

Confirm that:

Improper seating of the flow sensor, expiratory valve set, or valve membrane can prevent successful calibration.

Expected outcome: The expiratory valve and membrane are properly assembled and seated.

8. Repeat the Calibration Correctly

With the patient disconnected:

A red X indicates unsuccessful calibration.

Expected outcome: Calibration completes with a checkmark.

If calibration passes, perform the tightness test and remaining required preoperational checks before returning the ventilator to service.

9. Substitute a Known-Good Flow Sensor

Install a compatible, known-good Hamilton flow sensor and repeat the calibration.

Use the correct calibration adapter when required.

Expected outcome:

Do not return the original sensor to use unless it passes inspection and calibration.

10. Substitute External Circuit Components

If the known-good sensor also fails, replace external components one at a time and repeat calibration after each change:

Hamilton’s specified sequence after persistent failure is to replace the flow sensor, expiratory valve membrane, and expiratory valve before requesting service.

Expected outcome: The defective external component is isolated.

11. Perform the Tightness Test and Functional Verification

After successful calibration:

The C6 requires flow-sensor calibration and a circuit tightness test after installing a new or reprocessed circuit component.

Expected outcome: The ventilator passes calibration, tightness testing, and functional verification without alarms.

If the Problem Persists

If calibration continues to fail with:

then common external causes have been ruled out. The problem may involve the internal differential-pressure measurement system, pneumatic pathway, sensor interface, or associated electronics.

The ventilator should be:

Do not perform internal pneumatic or board-level repairs without the appropriate Hamilton service documentation, training, and test equipment. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never calibrate a flow sensor while the ventilator is connected to an active patient. Move the patient to another verified ventilator before disconnecting the circuit, replacing components, or running calibration tests.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported that the Hamilton C6 displayed a red X during flow-sensor calibration and would not complete the preoperational check."

Cause

What was observed during troubleshooting.

Example:
"Inspection found condensate inside the adult/pediatric flow sensor and moisture within one flow-sensor pressure tube."

Resolution

What action was taken.

Example:
"Replaced the flow sensor and affected tubing, successfully completed flow-sensor calibration and circuit tightness testing, verified operation with a test lung, and returned the ventilator to service."

Helpful Details to Include (If Known)

Final Thought

Flow-sensor calibration failures should be approached by protecting the patient first, then checking sensor selection, tubing, leaks, moisture, and replaceable circuit components. Escalate when known-good external components do not correct the failure, and document each finding clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide