Hamilton C6

O₂ Sensor Calibration Failure or Incorrect FiO₂

On this page

Asset Type

Ventilator

Manufacturer

Hamilton

Model

C6

What This Guide Helps With

Troubleshooting failed O₂ sensor calibration or inaccurate displayed FiO₂ caused by gas supply, connections, calibration conditions, settings, or sensor failure.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot unreliable oxygen delivery or measurement while the HAMILTON-C6 is actively supporting a patient.

Expected outcome: The patient is safely supported without relying on uncertain FiO₂ delivery or monitoring.

Continue Clinical Engineering troubleshooting only after the ventilator has been removed from patient use.

2. Confirm the Exact Failure

Review the displayed alarm, message, and alarm history.

Determine whether:

Record the set FiO₂, displayed oxygen value, gas-source status, and exact message.

Expected outcome: The failure pattern is clearly identified before components are changed.

3. Verify the Oxygen Supply

Confirm that the oxygen hose is:

Verify the oxygen source has adequate pressure and is not empty, shut off, or intermittently failing. When permitted, test the outlet using an approved gas-pressure analyzer or a known-good device.

An inadequate oxygen source can cause both calibration failure and incorrect delivered FiO₂.

Expected outcome: A stable oxygen supply is available at the ventilator inlet.

If correcting the gas supply resolves the issue, repeat the required preoperational checks and stop troubleshooting.

4. Check the Air Supply

Verify that the compressed-air connection is also correct and stable.

Incorrect air pressure, a crossed hose, or loss of one gas source can affect the ventilator’s oxygen blending and make the displayed or measured FiO₂ appear incorrect.

Expected outcome: Both required gas sources are correctly connected and available.

5. Inspect External Gas Connections

Inspect the oxygen and air fittings for:

Substitute known-good, compatible gas hoses when available.

Expected outcome: The ventilator receives gas through secure, undamaged connections.

If replacing a defective hose or fitting restores normal operation, complete operational verification and stop.

6. Verify the Oxygen Setting and Alarm Limits

Confirm that the selected oxygen concentration matches the intended test value.

Review:

Ensure the complaint is not caused by comparing the measured oxygen value with an outdated prescription, previous patient setup, or automatically adjusted setting.

Expected outcome: The intended oxygen setting and alarm configuration are confirmed.

7. Allow Adequate Stabilization Time

After connecting gas supplies or changing the oxygen setting, operate the ventilator on an approved test lung and allow the oxygen concentration to stabilize.

A brief difference between the set and displayed value may occur immediately after startup or a major FiO₂ adjustment. Persistent or excessive deviation requires further evaluation.

Expected outcome: The displayed oxygen value stabilizes near the selected setting.

8. Prepare the Ventilator for O₂ Sensor Calibration

Place the ventilator in the condition required by its installed software and sensor configuration. Follow the on-screen calibration instructions exactly.

Before starting:

The HAMILTON-C6 includes an O₂ sensor calibration function and can generate an O₂ sensor calibration needed alarm when calibration is required.

Expected outcome: The ventilator is in a stable and appropriate condition for calibration.

9. Run the O₂ Sensor Calibration

Initiate the O₂ sensor calibration from the appropriate System, Tests & Calib, or preoperational-check screen for the installed software version.

Follow all displayed prompts without disconnecting gas supplies or changing settings during calibration.

Expected outcome: Calibration completes successfully and the calibration-needed message clears.

If calibration succeeds, continue to independent FiO₂ verification before returning the ventilator to service.

10. Repeat Calibration After Correcting External Conditions

If calibration fails:

Do not repeatedly calibrate a sensor without correcting the condition causing the failure.

Expected outcome: A temporary connection, pressure, or stabilization issue is corrected.

11. Inspect the Accessible O₂ Sensor Area

With the ventilator powered down and removed from service, inspect only externally accessible sensor-related areas permitted by facility procedure.

Look for:

Do not perform deep internal disassembly or board-level troubleshooting.

Expected outcome: No external sensor seating or contamination problem remains.

12. Compare Delivered FiO₂ With an Independent Analyzer

Connect the ventilator to a suitable test lung and calibrated oxygen analyzer.

Test several clinically relevant oxygen settings, such as:

Allow the reading to stabilize at each setting. Compare the independent analyzer reading with the set and displayed FiO₂.

Use the manufacturer’s applicable specifications and facility test procedure to determine acceptance.

Expected outcome: Delivered oxygen tracks the selected concentration within the permitted tolerance.

If the external analyzer agrees with the set value but the ventilator display does not, suspect an O₂ sensing or calibration problem.

If both the external analyzer and ventilator indicate incorrect oxygen delivery, suspect a gas-source, mixer, valve, or internal pneumatic problem.

13. Test With a Known-Good Approved O₂ Sensor When Authorized

When the installed sensor is a replaceable item and facility policy permits, substitute a known-good Hamilton-approved compatible O₂ sensor.

Repeat:

Do not install an unapproved or incompatible sensor.

Expected outcome: Calibration and FiO₂ accuracy return to normal with the known-good sensor.

If replacement resolves the issue, document the failed sensor and complete all required functional testing before return to service.

14. Perform the Complete Preoperational Check

After any adjustment, hose substitution, calibration, or sensor replacement, perform the complete manufacturer-required preoperational check.

Verify:

Expected outcome: All required tests pass without alarms or abnormal oxygen readings.

If any required check fails, do not return the ventilator to clinical use.

If the Problem Persists

If gas sources, hoses, settings, circuit conditions, calibration preparation, and the accessible O₂ sensor have been evaluated, the common external causes have been ruled out.

The problem may involve:

The ventilator should be:

Knowing when to stop external troubleshooting and escalate the device is proper Clinical Engineering practice.

Clinical Use Tip

Never rely solely on the ventilator’s displayed FiO₂ when oxygen delivery is in question. Remove the patient from the affected ventilator and verify oxygen concentration with an independent calibrated analyzer before returning the unit to service.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported that the HAMILTON-C6 failed O₂ sensor calibration and displayed an FiO₂ value lower than the selected oxygen setting."

Cause

What was observed during troubleshooting.

Example:
"Verified stable wall air and oxygen supplies, then found that the installed O₂ sensor continued to fail calibration and produced unstable readings during independent oxygen testing."

Resolution

What action was taken.

Example:
"Removed the ventilator from service, replaced the approved O₂ sensor, successfully completed calibration and preoperational testing, and verified delivered FiO₂ with a calibrated oxygen analyzer."

Helpful Details to Include (If Known)

Final Thought

Oxygen-related faults require careful patient-safety decisions and objective testing. Verify external gas sources and connections first, follow the calibration prompts, confirm delivered FiO₂ independently, and escalate when calibration or accuracy cannot be restored. Complete CCR documentation provides a clear record of the complaint, findings, corrective action, and final equipment status.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide