Hamilton G5 Ventilator

Flow Sensor Errors

On this page

Asset Type

Ventilator

Manufacturer

Hamilton

Model

G5

What This Guide Helps With

This guide addresses situations where the Hamilton G5 ventilator displays flow sensor-related alarms or errors, such as:

The focus is on logical, external checks first — particularly sensor condition, circuit setup, moisture, and calibration — before assuming internal failure.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Identify the Exact Alarm Message

This helps determine whether the issue is setup-related or hardware-related.

3. Inspect the Flow Sensor for Moisture or Contamination

The Hamilton G5 flow sensor is highly sensitive to:

Moisture is one of the most common causes of flow sensor alarms.

4. Verify Proper Sensor Orientation and Connection

Improper orientation or loose connections can generate false readings or calibration failures.

5. Replace the Breathing Circuit

A compromised circuit can cause:

6. Perform Flow Sensor Calibration

If calibration fails repeatedly with multiple sensors and circuits, the issue may be internal.

7. Check for Excessive Condensation in the Expiratory Path

Inspect:

Drain any water traps. Excess moisture can affect measurement accuracy and trigger alarms.

8. Inspect for Physical Damage

Replace if any damage is present.

9. Power Cycle the Ventilator

Intermittent software-related sensor detection issues may clear with a restart.

If the Problem Persists

If multiple known-good sensors fail, multiple circuits fail, calibration repeatedly fails, or the alarm persists after moisture elimination, the likely cause may be:

At this point:

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Flow measurement accuracy directly affects tidal volume delivery and alarm safety.

Never attempt to “silence and monitor” a flow sensor error on an active patient. Always transition to a backup ventilator before troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Nursing reported ‘Flow Sensor Error’ alarm on Hamilton G5. Ventilator failed pre-use calibration.”

Cause

What was observed during troubleshooting.

Example:
“Flow sensor found with internal condensation. Circuit also contained excessive moisture.”

Resolution

What action was taken.

Example:
“Replaced flow sensor and breathing circuit. Performed successful flow sensor calibration. Ventilator passed operational check and returned to service.”

Helpful Details to Include

Final Thought

Flow sensor errors are often caused by moisture, contamination, or circuit issues rather than internal hardware failure. A systematic approach — dry, inspect, replace, calibrate — resolves most problems. When repeated calibration failures occur with known-good components, escalation is appropriate and protects patient safety. Thorough documentation supports trend tracking and equipment reliability.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide