Hamilton G5 Ventilator

Oxygen Blending / FiO₂ Mismatch

On this page

Asset Type

Ventilator

Manufacturer

Hamilton

Model

G5

What This Guide Helps With

This guide addresses situations where the Hamilton G5 ventilator:

The focus is on ruling out external and easily verifiable causes first—gas supply, connections, setup, and configuration—before assuming internal mixer or sensor failure.

Step-by-Step Troubleshooting

1. Confirm Patient Safety First

2. Verify Wall Gas Supply

Why this matters:
The G5 blends oxygen and medical air internally. If one source is absent, low, or cross-connected, the delivered FiO₂ will not match the set value.

3. Inspect Gas Hoses and DISS Fittings

4. Check for Crossed Gas Lines

5. Run the Pre-Use Check / Calibration

6. Verify Oxygen Sensor Status

7. Compare with External Oxygen Analyzer (If Available)

8. Check for Active Alarms or Messages

9. Observe Stability Over Time

If the Problem Persists

If:

Then the issue is likely internal to the ventilator (e.g., proportional solenoid valves, internal mixer, or flow control components).

Knowing when to stop prevents unnecessary risk and avoids inappropriate internal disassembly.

Clinical Use Tip

Never troubleshoot oxygen blending on an actively ventilated patient.

Always transition the patient to another ventilator before testing gas supply or performing calibrations. Oxygen delivery errors directly affect patient safety and can cause rapid desaturation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Respiratory therapist reported that the ventilator was set to 40% FiO₂ but displayed 55% and triggered high oxygen alarm.”

Cause

What was observed during troubleshooting.

Example:
“Air hose was partially kinked behind the equipment boom, restricting air supply and causing higher-than-set oxygen concentration.”

Resolution

What action was taken.

Example:
“Replaced air hose, verified supply pressures, recalibrated oxygen sensor, and confirmed FiO₂ accuracy with external analyzer. Ventilator passed system check and returned to service.”

Helpful Details to Include (If Known)

Final Thought

Oxygen blending issues should always be approached logically: verify supply first, confirm measurement second, and escalate only after external causes are ruled out. Patient safety depends on accurate FiO₂ delivery, and disciplined troubleshooting protects both the patient and the equipment.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide