On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
This guide assists Clinical Engineering in troubleshooting oxygen sensor errors or inaccurate FiO₂ readings on a Hamilton T1 ventilator. These issues are commonly caused by calibration problems, sensor wear, environmental factors, or setup issues before assuming internal failure.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Do not troubleshoot while the ventilator is connected to a patient.
- Move the patient to a confirmed working ventilator before proceeding.
Verify the Reported Issue
- Confirm the exact alarm or symptom (e.g., FiO₂ mismatch, sensor error, drift).
- Compare displayed FiO₂ with expected settings or an external oxygen analyzer if available.
- This helps determine if the issue is true inaccuracy or a display/calibration issue.
Check Oxygen Supply Source
- Ensure oxygen supply is properly connected and turned on.
- Verify adequate pressure from wall supply or cylinder.
- Low or inconsistent supply can cause inaccurate readings.
Inspect Gas Hoses and Connections
- Check for kinks, loose fittings, or leaks in the oxygen supply line.
- Ensure proper connection to the ventilator inlet.
- Leaks or restrictions can affect oxygen delivery and sensor readings.
Perform Oxygen Sensor Calibration
- Run the ventilator’s oxygen calibration procedure per standard operation.
- Ensure calibration is done in the correct environment (typically room air or 100% O₂ as required).
- Improper or skipped calibration is a common cause of inaccuracy.
Verify Ambient Conditions
- Ensure the ventilator is not exposed to extreme humidity, temperature, or contaminants.
- Oxygen sensors can drift if environmental conditions are outside normal operating range.
Inspect the Oxygen Sensor
- Check for expiration date (many O₂ sensors have a limited lifespan).
- Look for signs of damage, corrosion, or moisture intrusion.
- Aging or contaminated sensors are a common failure point.
Replace the Oxygen Sensor (if applicable)
- If calibration fails or readings remain inaccurate, replace the sensor with a known good one.
- Recalibrate after replacement.
- This is a standard field-level corrective action.
Check for Condensation or Moisture
- Inspect internal tubing and sensor area (externally accessible areas only).
- Excess moisture can affect sensor accuracy.
- Allow system to dry if needed.
Restart the Ventilator
- Power cycle the unit and allow it to complete self-test.
- Recheck oxygen readings after restart.
- This can clear temporary faults or calibration glitches.
Run Pre-Use Check
- Perform a full system check to verify proper operation.
- Ensure no additional gas delivery or sensor-related errors are present.
If the Problem Persists
If oxygen sensor errors or inaccurate readings continue after calibration, sensor replacement, and verification of gas supply:
The issue is likely internal (e.g., gas mixing system, internal circuitry, or calibration control fault).
- Remove the device from service
- Label it Out of Service
- Send for repair or bench evaluation
Knowing when to stop prevents unnecessary risk and is part of proper troubleshooting.
Clinical Use Tip
Do not troubleshoot oxygen accuracy issues on a ventilator connected to a patient. Always transition the patient to a confirmed working ventilator before testing or calibration to ensure safe and accurate oxygen delivery.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
“Ventilator displaying incorrect FiO₂ values and intermittent oxygen sensor error alarm.”
Cause
What was observed during troubleshooting.
Example:
“Oxygen sensor failed calibration and was found to be expired.”
Resolution
What action was taken.
Example:
“Replaced oxygen sensor, performed calibration, and verified accurate FiO₂ readings with external analyzer. Device passed pre-use check and returned to service.”
Helpful Details to Include
- Oxygen source verified (wall or cylinder)
- Supply pressure confirmed
- Calibration success or failure
- Sensor expiration date checked
- Sensor replaced (if applicable)
- External analyzer comparison (if available)
- Presence of moisture or condensation
- Alarm messages observed
- Final device status
Final Thought
Oxygen accuracy is critical for patient safety, and even small deviations can have clinical impact. Always start with simple checks like supply, calibration, and sensor condition before escalating. Recognizing when the issue moves beyond external causes ensures safe, efficient troubleshooting and proper device handling.
That is successful troubleshooting.