HeartSine Samaritan PAD Series

Self-Test Failure on Power-On

On this page

Asset Type

Defibrillator

Manufacturer

HeartSine

Model

Samaritan PAD Series

What This Guide Helps With

Troubleshooting self-test failure during power-on, typically caused by battery, PAD-Pak, electrode, or environmental issues preventing the AED from entering ready status.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Patient care continues safely with a functional device

Why it matters: A failed self-test means the AED cannot guarantee shock delivery

Verify the Self-Test Failure Indication

Expected: Failure is clearly indicated (no ready status or error alert)

Why it matters: Confirms this is a true failure, not a misinterpretation

Check PAD-Pak Installation (Most Common Cause)

Expected: AED recognizes PAD-Pak and may pass self-test after reseating

Why it matters: Improper installation prevents the device from completing checks

Check PAD-Pak Expiration and Type

Expected: Device completes self-test with a valid PAD-Pak

Why it matters: Expired or incompatible PAD-Paks will trigger failure

Inspect Electrode Connection (If Applicable)

Expected: No physical defects in electrode system

Why it matters: AED checks electrode integrity during self-test

Check for Environmental Conditions

Expected: Device operates within normal environmental range

Why it matters: Extreme conditions can cause temporary self-test failures

Power Cycle the AED

Expected: Device may recover from a transient fault

Why it matters: Clears temporary logic or sensor errors

Swap with Known Good PAD-Pak (Definitive External Test)

Expected:

Why it matters: This isolates the most common external failure point

If the Problem Persists

If all external causes have been ruled out, the failure is likely internal (capacitor, internal circuitry, or system diagnostics failure).

Remove the AED from service
Label as Out of Service
Send for repair or bench evaluation

Knowing when to stop prevents unnecessary risk and wasted time.

Clinical Use Tip

Never troubleshoot or attempt to validate an AED during an active patient emergency. Always switch to a known working AED first, then troubleshoot afterward.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"AED failed self-test on power-on and did not show ready status."

Cause

What was observed during troubleshooting.

Example:
"Failed PAD-Pak identified during troubleshooting."

Resolution

What action was taken.

Example:
"Replaced PAD-Pak with in-date unit; AED passed self-test and returned to service."

Helpful Details to Include (If Known)

Final Thought

Self-test failures on AEDs must be taken seriously, as they directly impact life-saving functionality. Focus on simple external causes first—especially the PAD-Pak—then escalate appropriately. Clear documentation ensures reliability tracking and patient safety.

That is successful troubleshooting.