HeartSine Samaritan PAD Series

Shock Delivery or Charging Failure

On this page

Asset Type

Defibrillator

Manufacturer

HeartSine

Model

Samaritan PAD Series

What This Guide Helps With

Failure to charge or deliver shock, including no charge tone, delayed charging, or shock button not activating due to pads, battery, or system faults.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Patient care is not interrupted
Why it matters: Shock failure is a critical event and requires immediate backup intervention

Verify the Reported Issue

Expected: Device should charge and prompt for shock when indicated
Why it matters: Confirms whether the issue is reproducible or intermittent

Check PAD-Pak Installation (Most Common Cause)

Expected: Device recognizes PAD-Pak and proceeds normally
Why it matters: The PAD-Pak supplies both energy and electrodes—improper seating prevents charging

Inspect Electrode Pad Condition and Connection

Expected: Pads are intact and recognized by the device
Why it matters: Poor pad condition or connection can prevent safe shock delivery

Check for Audible and Visual Prompts

Expected: Clear transition from analysis → charging → shock prompt
Why it matters: Missing steps indicate where the failure is occurring

Verify Battery Capability (PAD-Pak Condition)

Expected: Device initiates charging with a fresh PAD-Pak
Why it matters: Low battery capacity prevents adequate energy buildup for shock delivery

Check Environmental and Storage Conditions

Expected: Device operates within normal conditions
Why it matters: Extreme environments can affect battery performance and internal charging circuits

Power Cycle and Reattempt Operation

Expected: Device proceeds through normal operation
Why it matters: Clears temporary faults or software glitches

If the Problem Persists

If PAD-Pak, pads, and external conditions have been verified and the device still fails to charge or deliver a shock:

Knowing when to stop prevents unnecessary risk and ensures patient safety.

Clinical Use Tip

Never troubleshoot a defibrillator while it is assigned to active patient care. Always replace it with a functioning unit before testing or evaluation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"AED failed to charge and did not deliver shock during readiness check."

Cause

What was observed during troubleshooting.

Example:
"Expired PAD-Pak with insufficient battery capacity prevented charging."

Resolution

What action was taken.

Example:
"Replaced PAD-Pak with new unit and verified successful charge and shock prompt during test."

Helpful Details to Include (If Known)

Final Thought

Shock delivery failures must be approached with urgency and a structured process. Always rule out pads and battery first, prioritize patient safety, and escalate appropriately when internal faults are suspected. Clear documentation ensures reliability and accountability.

That is successful troubleshooting.