Failure to charge or deliver shock, including no charge tone, delayed charging, or shock button not activating due to pads, battery, or system faults.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- If the device is in use during an emergency and fails to deliver therapy, immediately switch to a backup defibrillator
- Continue CPR per protocol
Expected: Patient care is not interrupted
Why it matters: Shock failure is a critical event and requires immediate backup intervention
Verify the Reported Issue
- Power on the AED and observe behavior during a simulated or test scenario (if safe to do so)
- Listen for charge tones
- Observe prompts related to charging or shock delivery
- Check if the shock button activates when prompted
Expected: Device should charge and prompt for shock when indicated
Why it matters: Confirms whether the issue is reproducible or intermittent
Check PAD-Pak Installation (Most Common Cause)
- Remove and reseat the PAD-Pak (battery + electrode cartridge)
- Ensure it is fully inserted and latched
- Check expiration date on the PAD-Pak
Expected: Device recognizes PAD-Pak and proceeds normally
Why it matters: The PAD-Pak supplies both energy and electrodes—improper seating prevents charging
Inspect Electrode Pad Condition and Connection
- Verify pads are properly connected (integrated with PAD-Pak)
- Check for damaged, dried, or expired pads
- Ensure pads are not previously used
Expected: Pads are intact and recognized by the device
Why it matters: Poor pad condition or connection can prevent safe shock delivery
Check for Audible and Visual Prompts
- Listen for prompts such as:
- “Analyzing” but no “Charging”
- “Check pads” or similar warnings
- Observe status indicator and any LED behavior
Expected: Clear transition from analysis → charging → shock prompt
Why it matters: Missing steps indicate where the failure is occurring
Verify Battery Capability (PAD-Pak Condition)
- Check expiration date
- Replace with a known-good PAD-Pak if available
Expected: Device initiates charging with a fresh PAD-Pak
Why it matters: Low battery capacity prevents adequate energy buildup for shock delivery
Check Environmental and Storage Conditions
- Ensure device is within operating temperature range
- Look for signs of moisture or contamination
Expected: Device operates within normal conditions
Why it matters: Extreme environments can affect battery performance and internal charging circuits
Power Cycle and Reattempt Operation
- Turn the unit off
- Wait a few seconds
- Power back on and repeat test
Expected: Device proceeds through normal operation
Why it matters: Clears temporary faults or software glitches
If the Problem Persists
If PAD-Pak, pads, and external conditions have been verified and the device still fails to charge or deliver a shock:
- The issue is likely internal (charging circuit, capacitor, or control system failure)
- Remove the device from service
- Label Out of Service
- Send for repair or bench evaluation
Knowing when to stop prevents unnecessary risk and ensures patient safety.
Clinical Use Tip
Never troubleshoot a defibrillator while it is assigned to active patient care. Always replace it with a functioning unit before testing or evaluation.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"AED failed to charge and did not deliver shock during readiness check."
Cause
What was observed during troubleshooting.
Example:
"Expired PAD-Pak with insufficient battery capacity prevented charging."
Resolution
What action was taken.
Example:
"Replaced PAD-Pak with new unit and verified successful charge and shock prompt during test."
Helpful Details to Include (If Known)
- PAD-Pak expiration date checked
- PAD-Pak reseated or replaced
- Status indicator condition (green/red/no light)
- Audible prompts during operation
- Environmental conditions (temperature, storage)
- Whether shock button activated
- Final device status after testing
Final Thought
Shock delivery failures must be approached with urgency and a structured process. Always rule out pads and battery first, prioritize patient safety, and escalate appropriately when internal faults are suspected. Clear documentation ensures reliability and accountability.
That is successful troubleshooting.