Hillrom CareAssist

Actuator Failure or Slow Movement (Hi-Lo, Backrest, Knee Gatch)

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

CareAssist

What This Guide Helps With

Addresses slow, uneven, or non-functioning bed movements due to power, control, or connection issues before assuming actuator failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify Bed Has Power

Check Battery Status

Inspect Hand Control and Side Rail Controls

Check for Bed Lockout or CPR Mode Activation

Inspect for Obstructions or Mechanical Binding

Listen for Actuator Noise During Operation

Check Actuator and Cable Connections

Test Individual Functions Separately

Cycle the Bed Power

If the Problem Persists

If all external checks have been completed and movement is still slow or non-functional, the issue is likely internal (actuator, control box, or wiring harness).

Remove the bed from service, label it Out of Service, and send it for repair or bench evaluation.

Recognizing when to stop prevents unnecessary risk and is part of proper troubleshooting.

Clinical Use Tip

Always move the patient to another bed before troubleshooting movement issues. Unexpected bed motion or failure during use can create serious safety risks.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Bed backrest and hi-lo functions moving slowly and intermittently unresponsive."

Cause

What was observed during troubleshooting.

Example:
"Loose actuator cable connection identified at control box causing inconsistent power to actuator."

Resolution

What action was taken.

Example:
"Reseated and secured actuator connection, tested all bed movements, and confirmed normal operation."

Helpful Details to Include (If Known)

Final Thought

Actuator issues often appear more severe than they are, but many are caused by simple power, control, or connection problems. A structured, external-first approach ensures patient safety while avoiding unnecessary part replacement. Proper escalation and clear documentation are key to effective Clinical Engineering support.

That is successful troubleshooting.