On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Error 0x11A3 indicating a stuck right-side pendant button causing unintended bed movement or loss of normal pendant control.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Remove the bed from active patient use if unintended movement is occurring
- Transfer the patient to a safe, functioning bed if necessary
- Ensure no risk of uncontrolled bed movement
Verify the Reported Error and Behavior
- Confirm that Error 0x11A3 is displayed on the bed interface
- Observe if the bed moves without input
- Check if a specific function is continuously active
- Determine if the issue occurs at startup or during use
Inspect the Right Pendant for Physical Obstruction
- Check for debris or dried fluids around buttons
- Press buttons to confirm proper movement and return
- Ensure no object is pressing against the pendant
Clean the Pendant Surface
- Use approved disinfectant wipes to clean the pendant
- Dry thoroughly after cleaning
- Re-test button responsiveness
Check Pendant Cable and Connection
- Inspect cable for cuts, pinching, or strain damage
- Ensure connector is fully seated
- Reseat connection if accessible
Power Cycle the Bed
- Turn off or unplug the bed
- Wait briefly before restoring power
- Restart and check if error clears
Isolate Pendant vs Bed Issue
- Swap with a known-good pendant if available
- Determine if the error follows the pendant or remains with the bed
If the Problem Persists
If Error 0x11A3 continues after all external checks:
- External causes have been ruled out
- Likely internal pendant failure or control circuit fault
- Remove the bed from service
- Label the device as Out of Service
- Replace the pendant or send for repair/bench evaluation
Clinical Use Tip
Never troubleshoot a bed with active patient use. Always transfer the patient to a safe device before evaluating control-related issues.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Bed displaying Error 0x11A3 with right pendant causing continuous or unresponsive control input."
Cause
What was observed during troubleshooting.
Example:
"Right pendant button found sticking due to internal failure."
Resolution
What action was taken.
Example:
"Inspected and cleaned pendant; issue persisted. Removed bed from service and replaced pendant."
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
Stuck control inputs are safety-critical issues. Always begin with simple external checks, but escalate quickly when faults persist. Proper documentation and timely removal from service ensure patient safety and effective repairs.
That is successful troubleshooting.