On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
This guide helps troubleshoot Error 0x1E02 indicating a SafeView control board fault, which can result in loss of SafeView functionality, monitoring features, or communication issues within the bed system.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Remove the bed from patient use if SafeView monitoring is required
- Transfer the patient to a functioning bed if needed
- Do not rely on monitoring features during troubleshooting
Verify the Reported Error and Behavior
- Power on the bed and confirm Error 0x1E02 is displayed
- Observe when the error appears (startup or during use)
- Check if SafeView features are unavailable or intermittent
Perform a Basic Power Reset
- Unplug the bed from AC power
- Disconnect battery if accessible without disassembly
- Wait 30–60 seconds
- Reconnect power and restart the bed
Check External Connections Related to SafeView
- Inspect visible cables and harnesses associated with SafeView
- Ensure all connectors are fully seated
- Look for pinched, damaged, or disconnected wiring
Inspect for Environmental or Physical Issues
- Check for signs of liquid intrusion
- Inspect for physical damage to panels or housings
- Look for debris or contamination affecting electronics
Isolate as Device-Level Failure
- If error persists after reset and external checks, suspect internal failure
- Confirm no external issues are contributing
- Recognize this as a likely control board fault
If the Problem Persists
If all external causes have been ruled out and the error remains, the SafeView control board is likely faulty.
- Remove the bed from service
- Label the device as Out of Service
- Send for repair or replace the SafeView control board
Clinical Use Tip
Never troubleshoot a bed with an active patient. Always transfer the patient to a safe, functioning bed before addressing system faults.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Bed displaying Error 0x1E02 and SafeView system not functioning."
Cause
What was observed during troubleshooting.
Example:
"Persistent SafeView control board fault after power reset and cable inspection."
Resolution
What action was taken.
Example:
"Removed bed from service and sent for SafeView control board replacement."
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
System-reported control board faults should be taken seriously. After ruling out power and connection issues, escalation ensures patient safety and efficient resolution.
That is successful troubleshooting.