On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Error 0x1E03 indicating SafeView light failure, affecting under-bed lighting visibility and fall-risk safety features.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Remove the bed from active patient use if SafeView lighting is required
- Transfer the patient to a safe, functioning bed if necessary
Verify the Reported Error and Behavior
- Power on the bed and confirm Error 0x1E03 is displayed
- Check if SafeView light is not illuminating, flickering, or intermittent
- Observe whether the error appears at startup or during operation
Perform a Bed Power Reset
- Unplug the bed from AC power
- Remove battery if externally accessible
- Wait 60 seconds before restoring power
Check External Connections to SafeView Lighting
- Inspect under-bed cabling for loose or disconnected connectors
- Look for pinched, damaged, or worn wiring
Inspect SafeView Light Assembly
- Check for cracked housings or visible physical damage
- Look for signs of fluid intrusion or contamination
- Verify light components appear intact
Reseat Accessible Connectors
- Disconnect and reconnect any accessible harnesses to the light system
- Ensure secure and proper connection
If the Problem Persists
If Error 0x1E03 continues after external checks:
- External causes have been ruled out
- Remove the bed from service
- Label the device as Out of Service
- Replace the SafeView light board or send for repair
Clinical Use Tip
Never troubleshoot on an active patient. Always move the patient to a backup device before addressing equipment issues.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Bed displaying Error 0x1E03 with SafeView light not functioning."
Cause
What was observed during troubleshooting.
Example:
"SafeView light board failure confirmed after reset and cable inspection did not resolve issue."
Resolution
What action was taken.
Example:
"Removed bed from service and replaced SafeView light board."
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
Error codes like 0x1E03 help identify subsystem failures quickly. Always rule out simple external causes first, but escalate to board replacement when necessary to maintain safety and reliability.
That is successful troubleshooting.