Hillrom Centrella Smart+

Error: 0x1E03 SafeView Light Board Fault

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

Centrella Smart+

What This Guide Helps With

Error 0x1E03 indicating SafeView light failure, affecting under-bed lighting visibility and fall-risk safety features.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Perform a Bed Power Reset

Check External Connections to SafeView Lighting

Inspect SafeView Light Assembly

Reseat Accessible Connectors

If the Problem Persists

If Error 0x1E03 continues after external checks:

Clinical Use Tip

Never troubleshoot on an active patient. Always move the patient to a backup device before addressing equipment issues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Bed displaying Error 0x1E03 with SafeView light not functioning."

Cause

What was observed during troubleshooting.

Example:
"SafeView light board failure confirmed after reset and cable inspection did not resolve issue."

Resolution

What action was taken.

Example:
"Removed bed from service and replaced SafeView light board."

Helpful Details to Include (If Known)

Final Thought

Error codes like 0x1E03 help identify subsystem failures quickly. Always rule out simple external causes first, but escalate to board replacement when necessary to maintain safety and reliability.

That is successful troubleshooting.