Hillrom Centrella Smart+

Error: 0x1E06 SafeView Communication Hardware Error

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

Centrella Smart+

What This Guide Helps With

This guide addresses Error 0x1E06 SafeView Communication Hardware Error, which indicates a failure in communication hardware within the SafeView system. This can result in loss of bed exit monitoring and SafeView functionality, increasing patient fall risk.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check External Cable Connections (Most Common Cause)

Inspect Cable Assembly Condition

Eliminate Environmental or Setup Causes

Reseat or Replace Cable Assembly

Stop if Resolved

If the Problem Persists

If Error 0x1E06 continues after cable inspection and replacement, external causes have been ruled out and an internal communication hardware failure is likely.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup bed before addressing SafeView-related faults to ensure continuous fall protection.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Bed displaying Error 0x1E06 SafeView Communication Hardware Error; SafeView system non-functional."

Cause

What was observed during troubleshooting.

Example:
"Fault traced to defective SafeView cable assembly interrupting communication."

Resolution

What action was taken.

Example:
"Replaced cable assembly, verified connections, tested SafeView functionality, returned bed to service."

Helpful Details to Include (If Known)

Final Thought

Communication errors should always be approached from the outside in—starting with cables and connections. Proper troubleshooting and escalation ensure SafeView functionality is restored safely and efficiently.

That is successful troubleshooting.