Hillrom Centrella Smart+

Error: 0x2001 Audio Codec Driver Failure

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

Centrella Smart+

What This Guide Helps With

This guide addresses Error 0x2001 Audio Codec Driver Failure, which indicates a failure in the bed’s audio system. This can result in loss of audible alarms, alerts, and feedback tones, creating a potential patient safety risk if critical alerts are not heard.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Perform a Power Cycle

Check External Connections and Accessories

Inspect SCB ↔ MCB Cable Connections

Check SOM Board Seating

Isolate to Internal Hardware

If the Problem Persists

If all external connections, power cycling, and board seating checks are completed and the error remains, the issue is likely internal.

Clinical Use Tip

Never troubleshoot a bed with disabled audio alarms while a patient is relying on it. Always move the patient to a backup device before continuing diagnostics.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Bed displaying Error 0x2001 with no audible alarms or tones."

Cause

What was observed during troubleshooting.

Example:
"Audio codec failure suspected after power cycle and cable reseating did not resolve issue."

Resolution

What action was taken.

Example:
"Reseated SCB connections and verified SOM seating. Error persisted. Bed removed from service and SCB replacement initiated."

Helpful Details to Include (If Known)

Final Thought

Audio system failures directly impact patient safety by eliminating critical alerts. Always rule out simple external causes first, then escalate quickly when alarms are compromised. Clear documentation ensures accountability and efficient repair handling.

That is successful troubleshooting.