Hillrom Compella

Head, Knee, Foot, or High-Low Actuator Will Not Move Under Load

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

Compella

What This Guide Helps With

Troubleshoot bed articulation that fails or stalls under load due to power, lockouts, obstructions, positioning, or mechanical resistance.

Step-by-Step Troubleshooting

1. Ensure Patient Safety and Stable Support

Stop operating an actuator that stalls, binds, or moves unpredictably.

Expected outcome: The patient is protected and the bed is stable for evaluation.

2. Confirm Which Motion Is Affected

Identify the specific articulation failure.

Expected outcome: The failed movement and operating conditions are documented.

3. Verify Power Availability

Inspect the bed's electrical power condition.

Expected outcome: Reliable electrical power is available.

4. Check Lockouts and Control Response

Inspect applicable articulation controls.

Expected outcome: The affected function receives a valid command through normal controls.

5. Inspect for External Obstructions

Examine accessible areas around the deck.

Expected outcome: No external obstruction prevents the intended movement.

6. Verify Load and Positioning Conditions

Determine whether the reported condition is related to loading.

Expected outcome: The bed configuration is appropriate and no external loading condition explains the malfunction.

7. Perform an Unoccupied Articulation Test

With the bed unoccupied, operate the affected function within the approved range.

Expected outcome: The affected section moves smoothly without binding or abnormal noise.

If movement works unoccupied but fails under an approved load, do not consider the problem resolved.

8. Complete Required Functional Verification

After correcting an external issue:

Expected outcome: The actuator passes the required functional verification without stalling, binding, or unexpected movement.

Troubleshooting can stop only after the reported load-related condition has been adequately verified.

If the Problem Persists

If power, controls, lockouts, obstructions, positioning, and loading conditions have been checked, common external causes have been ruled out.

Possible service-level causes include actuator drive components, mechanical linkages, load-related resistance, motor control, position sensing, or associated wiring.

The bed should be:

Do not return a bed to service based only on successful unloaded movement when the reported failure occurred under load.

Clinical Use Tip

A bed that cannot reliably adjust height or patient position can create transfer, fall, and caregiver injury risks even when the mattress remains supportive.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the Compella head section would not raise while the patient was positioned in the bed."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found an attached accessory cable trapped in the head articulation path, restricting movement."

Resolution

What action was taken.

Example:
"Removed the obstruction and verified smooth head articulation and normal control operation under the approved functional test conditions."

Helpful Details to Include (If Known)

Final Thought

Articulation problems require careful attention to patient support, mechanical clearance, and loading. Verify external conditions before suspecting actuator failure, avoid forcing stalled mechanisms, escalate unresolved faults, and document successful return-to-service testing.

That is successful troubleshooting.