Hillrom Compella

Siderail Controls, Patient Pendant, or Lockout Function Is Unresponsive

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

Compella

What This Guide Helps With

Troubleshoot unresponsive siderail controls, patient pendant functions, or lockouts caused by power, connections, damaged controls, or incorrect settings.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Safe Positioning

Confirm that the patient can be positioned safely without relying on the affected controls.

Expected outcome: Patient positioning and fall-prevention needs are maintained while the controls are evaluated.

2. Confirm the Exact Control Failure

Determine which controls are affected.

Expected outcome: The affected control location and function are identified.

3. Verify Electrical Power

Check the bed's external power condition.

Expected outcome: The bed has stable power and unaffected functions operate normally.

4. Inspect the Patient Pendant and Cable

If equipped with the affected pendant:

Do not open sealed control assemblies.

Expected outcome: The pendant and external connection are intact and properly seated.

5. Inspect Siderail Control Panels

Examine the affected siderail controls.

Expected outcome: The control panel is physically intact and responds consistently.

6. Verify Lockout Status

Check the applicable lockout controls.

Expected outcome: Lockouts function correctly and only the intended controls are disabled.

7. Compare Available Control Locations

With the bed unoccupied:

Expected outcome: The problem is isolated to an external control or the bed control system.

8. Verify Control and Lockout Operation

After correcting an external issue:

Expected outcome: Controls and lockouts operate consistently and safely.

Troubleshooting can stop when all affected functions pass.

If the Problem Persists

If power, external connections, control condition, siderail positioning, and lockouts have been checked, common external causes have been ruled out.

Possible service-level causes include a control assembly, wiring, communication between control locations, or the bed control system.

The bed should be:

An unreliable lockout function must be treated as a safety concern.

Clinical Use Tip

Verify that caregiver lockouts prevent unintended patient operation while still allowing the functions required for safe clinical care.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the Compella patient pendant would not operate the head section."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the pendant connector partially disengaged from its accessible connection point."

Resolution

What action was taken.

Example:
"Secured the connector and verified head movement, pendant response, and normal lockout operation."

Helpful Details to Include (If Known)

Final Thought

Reliable patient and caregiver controls are essential for safe bed positioning. Check power, connections, physical control condition, and lockouts before assuming internal failure. Escalate unresolved faults and document the completed functional verification.

That is successful troubleshooting.