Hillrom Progressa

Bed Exit Alarm False Alarms or Will Not Arm

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

Progressa

What This Guide Helps With

Troubleshooting bed-exit alarms that activate unexpectedly, fail to arm, or do not respond because of positioning, setup, accessories, power, or sensing problems.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot an unreliable bed-exit alarm while depending on it as the patient’s primary fall-prevention measure.

Expected outcome: The patient remains protected without relying on an alarm that may not function correctly.

2. Confirm the Reported Behavior

Determine whether:

Record any displayed message, selected alarm mode, indicator status, and audible behavior.

Expected outcome: The failure is clearly identified and can be reproduced safely.

3. Confirm AC Power and Bed Status

Verify that:

Bed-exit monitoring depends on the bed’s weighing and sensing system. An unstable power condition or broader scale fault may prevent reliable operation.

Expected outcome: The bed powers normally without related fault messages.

If restoring power resolves the issue, verify alarm operation and stop.

4. Confirm the Patient Is Properly Positioned

Ensure the patient is:

Hillrom instructs users to center the patient and align the patient with the hip indicator before arming the bed-exit alarm.

Expected outcome: The patient’s weight is distributed normally across the bed’s sensing system.

If repositioning resolves the issue, confirm proper alarm operation and stop.

5. Verify the Correct Bed-Exit Mode

From the Graphical Caregiver Interface:

The Position mode is the most sensitive and may alarm during normal repositioning or sitting up. Exiting and Out of Bed modes allow progressively more movement before activating.

Expected outcome: The selected mode matches the patient’s mobility and clinical fall-risk plan.

If selecting the appropriate mode resolves the false alarms, verify operation and stop.

6. Check for Weight Supported Outside the Bed

Inspect for anything that may transfer weight onto or away from the bed, including:

Expected outcome: Patient weight and approved bed-mounted accessories are supported consistently by the bed frame.

If removing the external load resolves the issue, verify operation and stop.

7. Inspect Bed-Mounted Accessories and Hanging Equipment

Check for changes made after the bed was last zeroed, such as:

Remove unnecessary items and ensure required accessories are mounted correctly without contacting external surfaces.

Expected outcome: The sensing system is not affected by unstable or externally supported weight.

8. Check the Mattress and Sleep Surface

Verify that:

Expected outcome: The mattress and frame can transmit patient weight normally without binding or interference.

9. Re-Zero the Bed When Clinically Appropriate

An incorrect or outdated bed zero may cause arming problems or false alarms.

Before zeroing:

On the GCI:

Hillrom specifies that the patient must be out of the bed during the zeroing process.

Expected outcome: The bed establishes a stable baseline and the alarm arms normally.

If re-zeroing resolves the issue, perform a controlled functional test and stop.

10. Perform a Controlled Functional Test

Test the alarm without relying on an active high-risk patient.

Expected outcome: Each selected mode arms, remains stable, activates appropriately, and resets normally.

11. Inspect for External Mechanical Interference

With the bed removed from patient use, inspect underneath and around the frame for:

Do not loosen load-cell hardware or perform internal adjustments without the approved Hillrom service procedure.

Expected outcome: No external condition is affecting frame movement or weight sensing.

12. Power-Cycle the Bed

When no patient is dependent on the bed:

Expected outcome: The GCI and sensing system restart normally without persistent alarm or scale faults.

If the alarm operates correctly after restart, complete repeated testing before returning the bed to service.

If the Problem Persists

If power, patient position, alarm mode, accessories, mattress placement, bed zero, and external interference have been ruled out, the problem may involve:

The bed should be:

Knowing when to stop and escalate is proper troubleshooting. Do not attempt load-cell adjustment, calibration, wiring repair, or internal component replacement without approved service documentation and required test equipment.

Clinical Use Tip

Do not use a malfunctioning bed-exit alarm as the sole protection for a patient at risk of falling. Provide another verified bed, direct observation, or an approved alternative before troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the Hillrom Progressa bed-exit alarm activated repeatedly while the patient remained centered in bed."

Cause

What was observed during troubleshooting.

Example:
"The bed had been zeroed before additional pumps and an oxygen cylinder were attached, causing an unstable weight baseline."

Resolution

What action was taken.

Example:
"Removed the patient, configured the bed with all required accessories, re-zeroed the scale, selected Exiting mode, and verified proper local and nurse-call alarm operation through repeated functional tests."

Helpful Details to Include (If Known)

Final Thought

Reliable bed-exit monitoring depends on stable power, correct patient positioning, appropriate sensitivity selection, proper bed configuration, and an accurate zero baseline. Protect the patient first, eliminate external causes logically, escalate suspected sensing failures appropriately, and document the exact findings and final disposition.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide