On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting incomplete Chair Egress positioning or ineffective stand assistance caused by controls, brakes, footboard placement, obstructions, surface operation, or power issues.
Step-by-Step Troubleshooting
1. Ensure Patient Safety First
Do not troubleshoot Chair Egress or Stand Assist movement while the bed is supporting an active patient.
- Notify the responsible clinical staff.
- Transfer the patient to another appropriate bed when continued care is required.
- Secure all patient lines, drainage devices, and accessories before moving the bed.
- Do not place anyone beneath or between moving frame sections.
- Use approved safe-patient-handling equipment when the patient cannot stand safely.
Expected outcome: The patient is safely removed from dependence on a bed with unreliable positioning.
If the bed moves unexpectedly, will not stop, or cannot safely support the patient, remove it from service immediately.
2. Clarify the Reported Failure
Determine whether:
- The Chair control does not respond.
- The bed enters FullChair position but will not continue into Chair Egress.
- The bed will not lower completely.
- The foot section will not retract or lower.
- The seat or leg surface will not deflate.
- The bed does not tilt forward or lower the knee section.
- The back surface does not inflate for egress assistance.
- Movement stops and an audible tone or message appears.
Record the complete displayed message, indicator behavior, and the last movement completed.
Expected outcome: The failure is narrowed to a specific portion of the Chair Egress sequence.
3. Verify the Bed Is Connected to Reliable AC Power
Confirm that:
- The power cord is fully connected.
- The outlet is operational and appropriate for medical equipment.
- The power cord and plug have no visible damage.
- The bed indicates that external power is present.
Test the outlet with an approved receptacle tester or another verified device according to facility policy.
Chair Egress requires several powered frame and surface movements. A weak battery or missing AC supply may prevent the sequence from completing.
Expected outcome: The bed operates from stable external power.
If restoring AC power resolves the issue, complete a functional test and stop.
4. Confirm the Brakes Are Set
Engage the bed brakes fully and verify that the brake indicator shows the expected condition.
The Progressa operating instructions require the brakes to be set before Chair Egress is activated.
Attempt the function again without a patient.
Expected outcome: The bed recognizes the brake condition and permits the Chair Egress sequence.
If setting the brakes resolves the issue, verify repeatable operation and stop.
5. Inspect the Footboard and Foot-End Area
Remove the footboard before completing Chair Egress. Store it securely and do not leave it flat where it could become contaminated or create a trip hazard.
Inspect for:
- A footboard still installed after the bed reaches FullChair.
- Equipment attached to the footboard.
- Linens caught in the retracting foot section.
- Cables, hoses, drainage bags, or accessories obstructing movement.
- Furniture or equipment blocking the foot end of the bed.
With the footboard installed, the bed may stop at FullChair and provide a tone rather than continuing into the full egress position.
Expected outcome: The foot section has adequate clearance to retract and lower.
If removing the obstruction resolves the issue, inspect for damage, complete the movement test, and stop.
6. Check the Controls and Lockout Settings
Inspect the Chair control on the caregiver siderail and any connected caregiver pendant.
Confirm that:
- The button is not physically damaged, stuck, or contaminated.
- The siderail is securely positioned.
- No articulation or control lockout is preventing operation.
- Other caregiver frame controls respond normally.
- The pendant connector, when equipped, is fully seated and undamaged.
Test the Chair control from another available caregiver control location.
Expected outcome: At least one caregiver control initiates and sustains the Chair sequence.
If one control works and another does not, the issue is likely isolated to the affected siderail control, pendant, cable, or connection. Remove the bed from service for repair.
7. Use the Correct Control Sequence
With the bed empty, brakes set, footboard removed, and the area clear:
- Press and continuously hold the Chair control.
- Allow the bed to move through the FullChair sequence.
- Continue following any on-screen instructions.
- After the surface-deflation indication or three-tone signal, continue holding or press and hold the Chair control as directed.
- Confirm that the bed lowers, tilts, retracts the foot section, and lowers the knee section.
- Hold the Chair control again when testing the optional back-inflation egress assistance.
The control must generally remain held during movement. Releasing it may stop the sequence as a safety feature.
Expected outcome: The complete Chair Egress sequence operates without interruption.
If correct control operation resolves the complaint, return the bed to a flat position, repeat the test, and stop.
8. Inspect the Mattress Surface and Connections
If the frame moves but the seat, leg, or back sections do not inflate or deflate correctly:
- Confirm the installed surface is compatible with the bed configuration.
- Inspect visible surface hoses and connectors for disconnection, kinking, pinching, or damage.
- Confirm the mattress is centered and properly seated on the frame.
- Check for linens or accessories compressing the surface hoses.
- Listen for the surface blower or compressor during the commanded movement.
- Note any surface-related messages or alarms.
On equipped Progressa beds, Chair Egress includes deflation of the seat and leg sections, with optional back-section inflation for egress assistance.
Expected outcome: The surface responds correctly during the positioning sequence.
If reseating an accessible external connection resolves the issue, repeat the complete function test and stop.
9. Check for Mechanical Obstruction or Uneven Movement
Observe the empty bed while commanding Chair Egress.
Look and listen for:
- One frame section moving while another remains stationary.
- Binding, jerking, or uneven movement.
- Grinding, clicking, or repeated actuator noise.
- Damaged or bent foot-section components.
- Objects trapped in the frame.
- A section reaching its travel limit before the sequence is complete.
- Movement that stops consistently at the same position.
Do not force any frame section or manually push the bed through the sequence.
Expected outcome: All frame sections move smoothly, evenly, and without unusual noise.
If movement is uneven, mechanically restricted, or accompanied by unusual noise, stop testing and remove the bed from service.
10. Perform a Controlled Power Reset
Only after the bed is empty and clinically released:
- Return the bed to a stable position when possible.
- Disconnect the bed from AC power.
- Follow facility-approved procedures to place the bed in a fully powered-down state.
- Wait briefly.
- Reconnect the bed to a verified AC outlet.
- Allow the system to initialize completely.
- Retest Chair Egress without a patient.
Do not repeatedly reset a bed that displays persistent errors or moves unpredictably.
Expected outcome: Temporary control or communication faults clear and normal operation returns.
If the function works after the reset, repeat the test several times and verify all related controls before returning the bed to service.
11. Complete a Final Functional and Safety Check
Before returning the bed to clinical use, verify:
- Brakes and steering operate correctly.
- Bed height adjustment functions normally.
- Head, knee, and foot sections move smoothly.
- FullChair and Chair Egress complete without interruption.
- Surface inflation and deflation operate as equipped.
- Both available caregiver controls function correctly.
- The bed returns safely to the flat position.
- No error messages, unusual sounds, or unintended movements occur.
- Power cord, siderails, footboard, and accessories are secure.
Expected outcome: The bed repeatedly completes the intended positioning sequence and is safe for clinical use.
If the Problem Persists
If the Progressa still will not complete Chair Egress or provide the expected egress assistance after power, brakes, controls, footboard placement, obstructions, surface connections, and operating sequence have been checked, common external causes have been ruled out.
The problem may involve an actuator, position sensor, brake-status input, control switch, surface-control component, wiring connection, or internal control system.
The bed should be:
- Removed from service.
- Labeled Out of Service.
- Placed in a safe, stable position.
- Sent for authorized repair or bench evaluation.
- Evaluated using the manufacturer’s service documentation and approved diagnostic equipment.
Do not bypass interlocks, force frame movement, or continue operating a bed that moves unpredictably. Knowing when to stop is proper troubleshooting.
Clinical Use Tip
Do not test Chair Egress or Stand Assist using a patient as the functional load. Transfer the patient first, secure the area, and test the bed empty. During clinical use, staff must monitor the patient, lines, drainage devices, linens, and foot placement throughout the egress sequence.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the Hillrom Progressa entered FullChair position but would not continue into Chair Egress or inflate the back section for stand assistance."
Cause
What was observed during troubleshooting.
Example:
"Inspection found the footboard remained installed and linens were obstructing the retracting foot section, preventing completion of the Chair Egress sequence."
Resolution
What action was taken.
Example:
"Removed the footboard and obstruction, verified brakes and AC power, and successfully tested FullChair, Chair Egress, surface deflation, back inflation, and return-to-flat operation."
Helpful Details to Include (If Known)
- Patient transferred before troubleshooting
- Exact error message or tone sequence
- AC outlet tested
- External power indicator confirmed
- Battery condition observed
- Brakes set and recognized
- Footboard removed
- Foot-end clearance verified
- Control lockouts checked
- Siderail and pendant controls tested
- Mattress and hose connections inspected
- Surface inflation or deflation observed
- Unusual sounds, binding, or uneven movement
- Power reset performed
- Complete Chair Egress sequence tested
- Final device status documented
Final Thought
Chair Egress failures should be approached as coordinated frame, control, and surface problems rather than immediately assuming actuator or internal-control failure. Protect the patient first, verify the simple prerequisites, observe exactly where the sequence stops, and escalate any unreliable or uneven movement. Clear CCR documentation helps the repair technician reproduce the failure and confirms why the bed was removed from service.
That is successful troubleshooting.