Hillrom Progressa

Chair Egress or Stand Assist Function Failure

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

Progressa

What This Guide Helps With

Troubleshooting incomplete Chair Egress positioning or ineffective stand assistance caused by controls, brakes, footboard placement, obstructions, surface operation, or power issues.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot Chair Egress or Stand Assist movement while the bed is supporting an active patient.

Expected outcome: The patient is safely removed from dependence on a bed with unreliable positioning.

If the bed moves unexpectedly, will not stop, or cannot safely support the patient, remove it from service immediately.

2. Clarify the Reported Failure

Determine whether:

Record the complete displayed message, indicator behavior, and the last movement completed.

Expected outcome: The failure is narrowed to a specific portion of the Chair Egress sequence.

3. Verify the Bed Is Connected to Reliable AC Power

Confirm that:

Test the outlet with an approved receptacle tester or another verified device according to facility policy.

Chair Egress requires several powered frame and surface movements. A weak battery or missing AC supply may prevent the sequence from completing.

Expected outcome: The bed operates from stable external power.

If restoring AC power resolves the issue, complete a functional test and stop.

4. Confirm the Brakes Are Set

Engage the bed brakes fully and verify that the brake indicator shows the expected condition.

The Progressa operating instructions require the brakes to be set before Chair Egress is activated.

Attempt the function again without a patient.

Expected outcome: The bed recognizes the brake condition and permits the Chair Egress sequence.

If setting the brakes resolves the issue, verify repeatable operation and stop.

5. Inspect the Footboard and Foot-End Area

Remove the footboard before completing Chair Egress. Store it securely and do not leave it flat where it could become contaminated or create a trip hazard.

Inspect for:

With the footboard installed, the bed may stop at FullChair and provide a tone rather than continuing into the full egress position.

Expected outcome: The foot section has adequate clearance to retract and lower.

If removing the obstruction resolves the issue, inspect for damage, complete the movement test, and stop.

6. Check the Controls and Lockout Settings

Inspect the Chair control on the caregiver siderail and any connected caregiver pendant.

Confirm that:

Test the Chair control from another available caregiver control location.

Expected outcome: At least one caregiver control initiates and sustains the Chair sequence.

If one control works and another does not, the issue is likely isolated to the affected siderail control, pendant, cable, or connection. Remove the bed from service for repair.

7. Use the Correct Control Sequence

With the bed empty, brakes set, footboard removed, and the area clear:

The control must generally remain held during movement. Releasing it may stop the sequence as a safety feature.

Expected outcome: The complete Chair Egress sequence operates without interruption.

If correct control operation resolves the complaint, return the bed to a flat position, repeat the test, and stop.

8. Inspect the Mattress Surface and Connections

If the frame moves but the seat, leg, or back sections do not inflate or deflate correctly:

On equipped Progressa beds, Chair Egress includes deflation of the seat and leg sections, with optional back-section inflation for egress assistance.

Expected outcome: The surface responds correctly during the positioning sequence.

If reseating an accessible external connection resolves the issue, repeat the complete function test and stop.

9. Check for Mechanical Obstruction or Uneven Movement

Observe the empty bed while commanding Chair Egress.

Look and listen for:

Do not force any frame section or manually push the bed through the sequence.

Expected outcome: All frame sections move smoothly, evenly, and without unusual noise.

If movement is uneven, mechanically restricted, or accompanied by unusual noise, stop testing and remove the bed from service.

10. Perform a Controlled Power Reset

Only after the bed is empty and clinically released:

Do not repeatedly reset a bed that displays persistent errors or moves unpredictably.

Expected outcome: Temporary control or communication faults clear and normal operation returns.

If the function works after the reset, repeat the test several times and verify all related controls before returning the bed to service.

11. Complete a Final Functional and Safety Check

Before returning the bed to clinical use, verify:

Expected outcome: The bed repeatedly completes the intended positioning sequence and is safe for clinical use.

If the Problem Persists

If the Progressa still will not complete Chair Egress or provide the expected egress assistance after power, brakes, controls, footboard placement, obstructions, surface connections, and operating sequence have been checked, common external causes have been ruled out.

The problem may involve an actuator, position sensor, brake-status input, control switch, surface-control component, wiring connection, or internal control system.

The bed should be:

Do not bypass interlocks, force frame movement, or continue operating a bed that moves unpredictably. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Do not test Chair Egress or Stand Assist using a patient as the functional load. Transfer the patient first, secure the area, and test the bed empty. During clinical use, staff must monitor the patient, lines, drainage devices, linens, and foot placement throughout the egress sequence.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the Hillrom Progressa entered FullChair position but would not continue into Chair Egress or inflate the back section for stand assistance."

Cause

What was observed during troubleshooting.

Example:
"Inspection found the footboard remained installed and linens were obstructing the retracting foot section, preventing completion of the Chair Egress sequence."

Resolution

What action was taken.

Example:
"Removed the footboard and obstruction, verified brakes and AC power, and successfully tested FullChair, Chair Egress, surface deflation, back inflation, and return-to-flat operation."

Helpful Details to Include (If Known)

Final Thought

Chair Egress failures should be approached as coordinated frame, control, and surface problems rather than immediately assuming actuator or internal-control failure. Protect the patient first, verify the simple prerequisites, observe exactly where the sequence stops, and escalate any unreliable or uneven movement. Clear CCR documentation helps the repair technician reproduce the failure and confirms why the bed was removed from service.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide