Hillrom Progressa

Surface Pressure Zone or Therapy Mode Fault

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

Progressa

What This Guide Helps With

Troubleshooting pressure-zone, inflation, Max-Inflate, rotation, percussion, vibration, or other surface therapy faults caused by setup, connections, positioning, or surface problems.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot an unreliable surface or therapy function while it is supporting an active patient.

Expected outcome: The patient is safely supported without relying on the affected surface.

If the surface cannot maintain appropriate inflation, remove the bed from clinical use.

2. Confirm the Exact Fault

Record the complete message displayed on the Graphical Caregiver Interface and determine whether the complaint involves:

The Progressa may be equipped with prevention, pulmonary, or therapy surface configurations, so available functions can vary by bed.

Expected outcome: The affected function and surface configuration are clearly identified.

3. Check AC Power and Bed Status

Verify that:

Surface inflation and powered therapy functions require reliable electrical power.

Expected outcome: The bed remains powered without interruptions or power-related messages.

If restoring AC power resolves the fault, perform a functional check and stop troubleshooting.

4. Allow the Surface to Complete Initialization

If the bed was recently powered on, transported, disconnected, or placed into CPR deflation:

Expected outcome: All surface zones inflate evenly and the selected surface mode becomes available.

If the surface initializes normally and maintains pressure, verify operation and stop.

5. Inspect the Surface Externally

Remove linens only as necessary and inspect for:

Do not probe punctures, patch the surface, or open the surface assembly unless specifically authorized by Hillrom service documentation.

Expected outcome: The surface is correctly positioned, unobstructed, and free of visible damage.

If visible damage is found, remove the bed from service and arrange surface replacement or repair.

6. Inspect Accessible Surface Connections

With the patient removed and the bed safely powered down when appropriate:

Do not remove covers or access internal pneumatic components.

Expected outcome: All accessible pneumatic connections are secure and undamaged.

If reseating an accessible connection resolves the fault, restart the bed and complete a functional test.

7. Confirm Bed and Surface Positioning

Place the bed in a safe, neutral position and verify:

Some therapy functions may be restricted by bed position, patient position, surface configuration, or an active safety condition.

Expected outcome: The bed permits the requested surface or therapy function without a positioning-related message.

8. Verify Patient Weight Information

Confirm that:

The Progressa surface uses bed and patient information to support pressure redistribution and therapy operation.

Expected outcome: The displayed weight is plausible and the surface pressure zones stabilize.

If correcting the weight setup resolves the issue, test the surface again and stop.

9. Test a Basic Surface Function

With no patient on the bed, or with an approved test load when required by facility procedures:

Expected outcome: The surface changes modes, inflates evenly, and maintains pressure without a fault.

If a specific zone remains deflated or the pump runs continuously, suspect a leak, valve problem, sensor problem, or internal pneumatic failure.

10. Test the Reported Therapy Mode

Only when safe and when the installed surface supports the function:

Progressa configurations may support continuous lateral rotation therapy and, on applicable surfaces, percussion and vibration therapy.

Expected outcome: The therapy starts, operates, and stops without a fault.

If the therapy repeatedly stops or faults under correct conditions, discontinue testing and escalate.

11. Perform a Controlled Restart

After the patient has been removed and clinical staff have approved:

Do not repeatedly power-cycle a bed that continues to display the same fault.

Expected outcome: The surface completes initialization and operates normally.

If the fault returns, treat it as a persistent failure.

12. Verify Operation Before Returning the Bed to Service

Before releasing the bed:

Expected outcome: The surface maintains pressure and all applicable therapy functions operate reliably.

If the Problem Persists

If the fault remains after external power, surface position, hose connections, patient-weight information, settings, and accessible surface components have been checked, common external causes have been ruled out.

The problem may involve an internal blower, valve manifold, pressure sensor, control board, wiring connection, pneumatic leak, or damaged surface component.

The bed should be:

Do not continue repeatedly operating a surface that cannot maintain pressure or reliably complete prescribed therapies. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a Progressa pressure-zone or therapy failure while a vulnerable patient remains dependent on the surface. Transfer the patient before testing inflation, disconnecting hoses, restarting the bed, or activating therapy modes.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the Progressa displayed a surface pressure-zone fault and would not start continuous lateral rotation therapy."

Cause

What was observed during troubleshooting.

Example:
"Inspection found the accessible surface air connector partially disengaged, preventing the surface from maintaining pressure in the affected zone."

Resolution

What action was taken.

Example:
"Removed the bed from patient use, reseated and secured the surface connector, restarted the bed, and verified normal inflation, Max-Inflate, and rotation therapy operation."

Helpful Details to Include (If Known)

Final Thought

Surface and therapy faults can directly affect pressure-injury prevention, pulmonary treatment, and patient safety. Begin with safe patient transfer and logical external checks, then escalate when the surface cannot maintain pressure or therapy operation. Clear CCR documentation helps the repair team reproduce the failure and return the bed to service safely.

That is successful troubleshooting.