On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting unresponsive touchscreen, siderail caregiver controls, or pendant functions caused by power, lockouts, connections, contamination, or external control damage.
Step-by-Step Troubleshooting
1. Ensure Patient Safety First
Do not troubleshoot bed positioning controls while the bed is supporting an unstable, unattended, or actively treated patient.
- Notify the responsible clinical staff.
- Transfer the patient to another verified bed when positioning, therapy, or emergency functions cannot be reliably operated.
- Confirm the brakes are engaged.
- Keep the bed in its lowest safe position whenever possible.
- Ensure lines, drains, tubes, and other equipment will not be pulled if the bed moves unexpectedly.
Expected outcome: The patient is protected from unintended bed movement, loss of positioning, or delayed access to required functions.
If essential positioning or emergency controls cannot be operated safely, remove the bed from clinical use.
2. Clarify Which Controls Are Unresponsive
Determine whether the problem affects:
- The Graphic Caregiver Interface touchscreen.
- One caregiver siderail control panel.
- Both caregiver siderail control panels.
- The patient pendant.
- Patient-side controls only.
- A single function, such as head, knee, height, chair, or Trendelenburg.
- All bed articulation and touchscreen functions.
Test only when the bed is unoccupied or when clinical staff confirm testing is safe.
Expected outcome: The affected control area and failure pattern are clearly identified.
A single failed control station suggests a localized control or connection problem. Multiple unresponsive control locations may indicate lockout, power, software, or internal communication failure.
3. Confirm the Bed Has Electrical Power
Verify that:
- The power cord is fully inserted into a known-good hospital-grade outlet.
- The outlet is energized.
- The power cord and plug show no cuts, bent blades, looseness, or heat damage.
- The bed’s power indicators illuminate.
- The bed responds when connected to another verified outlet.
Do not use an extension cord or unapproved power strip.
Expected outcome: The bed powers normally from a verified AC source.
If operation returns after correcting the outlet or power connection, confirm all controls function normally and stop troubleshooting.
4. Allow the Bed to Recharge
If the bed was unplugged for an extended period, connect it to AC power and allow the battery system to charge.
Observe whether:
- The touchscreen becomes active.
- Control indicators illuminate.
- Bed functions respond while connected to AC power.
- Battery or low-power warnings clear.
A severely discharged battery may prevent normal operation when the bed is unplugged.
Expected outcome: Controls operate normally while the bed is connected to AC power.
If controls work on AC power but fail immediately when unplugged, remove the bed from service for battery evaluation.
5. Check for Activated Control Lockouts
Inspect the caregiver control panel and touchscreen for illuminated lockout indicators.
Verify whether articulation functions have been intentionally locked out, including:
- Head movement.
- Knee movement.
- Bed height.
- Patient controls.
- Other configured functions.
The Progressa caregiver controls include lockout functions that can disable selected articulation controls.
Release only those lockouts that clinical staff confirm should be enabled.
Expected outcome: Previously locked functions respond normally after the appropriate lockout is released.
If releasing the lockout restores operation, verify the control from each applicable location and stop troubleshooting.
6. Confirm the Screen Is Not Displaying a Blocking Message
Inspect the touchscreen for:
- Warning messages.
- Help screens.
- Confirmation prompts.
- Service indicators.
- Active therapy screens.
- A screen that is illuminated but frozen.
- An unusually dim display.
Acknowledge only messages that can be safely cleared without changing prescribed patient therapy or clinical settings.
Record the exact wording of any persistent message.
Expected outcome: The normal caregiver interface is displayed and accepts input.
7. Clean and Dry the Control Surfaces
Inspect the touchscreen, siderail controls, and pendant for:
- Moisture.
- Cleaning solution residue.
- Gel or lotion.
- Adhesive residue.
- Linens pressing against buttons.
- Dirt around control edges.
- Physical objects contacting the touchscreen.
Clean the external surfaces using facility-approved methods, then dry them completely.
Do not spray liquid directly into control panels, connectors, or seams.
Expected outcome: Controls respond normally after external contamination or unintended contact is removed.
If cleaning resolves the issue, test each affected control several times and stop troubleshooting.
8. Inspect the Pendant and External Cable
If the pendant is affected:
- Examine the pendant housing for cracks, fluid intrusion, or impact damage.
- Inspect the cable for cuts, crushing, stretching, or sharp bends.
- Check the connector for looseness or visible damage.
- Confirm the cable is not trapped beneath the bed frame, siderail, caster, or mattress.
- Reseat only user-accessible connectors permitted by facility procedure.
Do not open the pendant housing or disassemble the connector.
Expected outcome: The pendant is securely connected and functions without intermittent response.
If a known-compatible, facility-approved pendant is available, substitution may help distinguish a pendant failure from a bed-side control problem.
9. Inspect the Siderails and Control Panels
Check each siderail for:
- Full mechanical engagement.
- Impact damage.
- Loose or cracked control overlays.
- Buttons that remain physically depressed.
- Evidence of fluid entry.
- Pinched or exposed wiring.
- Failure limited to a siderail that was recently raised, lowered, or struck.
Raise and lower the siderail only when safe, then confirm it locks securely.
Expected outcome: The siderail is mechanically secure and its controls operate normally.
A control panel that works intermittently as the siderail moves should be considered unreliable. Remove the bed from service.
10. Perform a Controlled Power Reset
Only perform a reset when the bed is unoccupied or the patient has been safely transferred.
- Place the bed in a stable, low position if controls permit.
- Disconnect the AC power cord.
- Allow the bed to remain disconnected briefly according to facility procedure.
- Reconnect the bed directly to a verified outlet.
- Allow the system to complete its startup process.
- Do not repeatedly press controls during startup.
Expected outcome: The touchscreen initializes and the caregiver controls respond normally.
If operation returns, complete a full functional check before returning the bed to service. A recurring freeze or need for repeated resets requires bench evaluation.
11. Test Each Control Location Separately
After power and lockouts are verified, test:
- Left caregiver siderail controls.
- Right caregiver siderail controls.
- Touchscreen functions.
- Pendant controls.
- Patient controls, when applicable.
- Head, knee, height, and other basic articulation functions.
The Progressa provides caregiver siderail controls for functions such as head and knee positioning, with position information also available through the caregiver interface.
Expected outcome: Each control produces the intended movement without delay, sticking, or intermittent operation.
Stop immediately if movement is unexpected, reversed, intermittent, or continues after the control is released.
12. Check for Evidence of Internal Failure
Remove the bed from service if any of the following remain:
- The touchscreen stays blank despite verified AC power.
- The screen freezes repeatedly.
- Multiple control stations remain unresponsive.
- Controls work intermittently.
- Bed movement occurs without a command.
- A function continues after the button is released.
- A control panel shows liquid damage.
- A persistent service code or communication error appears.
- There is unusual heat, odor, sound, or visible damage.
Do not open the touchscreen, siderails, control modules, or electrical enclosures unless trained and authorized under the manufacturer’s service procedures.
Hillrom provides dedicated technical training and service documentation for Progressa maintenance and diagnostic work beyond basic external troubleshooting.
If the Problem Persists
The common external causes—patient-control lockouts, AC power, discharged batteries, contamination, pendant connections, siderail condition, and temporary software state—have been ruled out.
The problem is likely related to an internal control panel, touchscreen assembly, communication circuit, power distribution component, wiring harness, or system software fault.
The bed should be:
- Removed from service.
- Labeled Out of Service.
- Sent for bench evaluation or qualified repair.
- Functionally and electrically tested before being returned to clinical use.
Knowing when to stop and escalate is proper troubleshooting. An intermittently responsive control system should not be considered safe for patient care.
Clinical Use Tip
Never test bed articulation on an active patient unless the responsible clinical staff have determined it is safe. Transfer the patient first when positioning controls are unreliable.
Remember that a bed with an unresponsive touchscreen may still have some mechanical controls available, but partial operation does not establish that the bed is safe for continued use.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the Progressa touchscreen and right caregiver siderail controls did not respond to commands."
Cause
What was observed during troubleshooting.
Example:
"Verified AC power and control lockouts; the right siderail control panel remained unresponsive and showed evidence of intermittent operation when the siderail moved."
Resolution
What action was taken.
Example:
"Removed the bed from service, labeled it Out of Service, and sent it for bench evaluation of the siderail control and communication system."
Helpful Details to Include (If Known)
- Patient transferred before troubleshooting
- Bed connected to a verified outlet
- Power cord and plug inspected
- Battery or power indicators observed
- Lockout indicators checked
- Exact touchscreen message recorded
- Left and right caregiver controls tested
- Pendant inspected or substituted
- Siderails checked for damage and secure engagement
- Touchscreen cleaned and dried
- Controlled power reset performed
- Intermittent or unintended movement observed
- Unusual sound, heat, odor, or fluid intrusion noted
- Final device status documented
Final Thought
Begin with patient safety, then isolate the failure by checking power, lockouts, contamination, accessories, and each control location. Escalate intermittent or persistent control failures and document exactly what was tested and found.
That is successful troubleshooting.