Hillrom TotalCare

Hi-Lo or Trendelenburg Function Failure

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

TotalCare

What This Guide Helps With

Identifies causes of bed height or tilt movement failure, including power, connection, actuator, and control issues before assuming internal motor or board failure.

Step-by-Step Troubleshooting

Ensure Patient Safety

Confirm the patient is safely transferred to another bed or positioned securely. Never attempt movement troubleshooting with an active patient in the bed.

Verify Power Supply

Check Battery Backup

Inspect Control Panel & Handset Connections

Test Movement Functions Individually

Check for Obstructions or Mechanical Locking

Swap Actuator Cables if Accessible

If the Problem Persists

External checks have been exhausted. The issue may involve internal actuator failure, motor controller, or mainboard. Bed should be:

Stopping here ensures safe and proper troubleshooting.

Clinical Use Tip

Never adjust hi-lo or Trendelenburg with a patient in bed. Always secure the patient in another bed or stretcher first to prevent injury.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Bed will not raise, lower, or tilt into Trendelenburg."

Cause

What was observed during troubleshooting.

Example:
"Handset and power verified; actuator shows no response when tested."

Resolution

What action was taken.

Example:
"Bed removed from service and sent to repair for actuator evaluation."

Helpful Details to Include (If Known)

Final Thought

Patient safety and methodical troubleshooting are critical. Starting with simple checks prevents unnecessary disassembly. Proper documentation ensures clear communication and supports repair follow-up.

That is successful troubleshooting.