Hillrom VersaCare

Error: Err0 Scale Display Error

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

VersaCare

What This Guide Helps With

This guide helps troubleshoot Err0 on the Hillrom VersaCare scale display when the bed scale system reports a scale control or scale board fault. This error may not stop normal bed movement, but it can affect patient weight accuracy and scale reliability.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Remove the Bed From Active Scale Use

Check External Power First

Cycle Bed Power

Press the Enable Key

If Err0 Clears, Perform Function Checks

Inspect for Obvious External or Physical Issues

Determine Whether the Issue Is Device-Related

Escalate for Scale P.C. Board Repair

If the Problem Persists

If Err0 remains after cycling power, waiting 10 seconds, reconnecting AC power, pressing Enable, and checking for obvious external issues, common user-level and external causes have been ruled out.

At that point, the condition strongly indicates an internal scale system failure, most likely involving the scale P.C. board or related scale electronics.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device first if troubleshooting requires movement, repair handling, or any action that could affect patient safety.

Do not rely on the VersaCare bed scale for active patient weight decisions while Err0 is present. Use another approved weighing method until the bed scale passes function checks.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Hillrom VersaCare bed reported with Err0 on the scale display. Scale function could not be verified for patient weight use."

Cause

What was observed during troubleshooting.

Example:
"Confirmed Err0 on scale display. AC power connection checked. Bed power cycled with 10-second wait before reconnecting to AC. Enable key pressed on siderail controls. Error remained, indicating likely internal scale electronics or scale P.C. board fault."

Resolution

What action was taken.

Example:
"Removed bed from service and labeled Out of Service. Sent for repair/bench evaluation. Scale P.C. board replacement recommended per troubleshooting path. If error remains after board replacement, escalate to Hill-Rom Technical Support."

Helpful Details to Include (If Known)

Final Thought

Err0 on a VersaCare scale display should be handled as a scale reliability issue until proven otherwise. Start with patient safety, verify the exact error, check power and controls first, then stop when the fault points beyond external troubleshooting. Clear documentation helps show what was checked, what was ruled out, and why the bed was returned to service or escalated for repair.

That is successful troubleshooting.