Hillrom VersaCare

Error: Err1 Scale Display Error

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

VersaCare

What This Guide Helps With

This guide helps troubleshoot Err1 on the Hillrom VersaCare scale display. Err1 indicates the bed scale system is not operating normally or is not clearing as expected. The bed may still move, but the scale should not be trusted for patient weight documentation until the issue is corrected and function checks pass.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error

Check AC Power and Basic Bed Power

Cycle Bed Power

Enable the Siderail Controls

Check for External Scale Interference

Recheck Scale Operation

Perform Function Checks if the Error Clears

Escalate if Err1 Remains

If the Problem Persists

If Err1 remains after basic power checks, a 10-second AC power cycle, siderail Enable activation, and external scale interference checks, common external causes have been ruled out.

Remove the device from service, label it Out of Service, and send it for repair or bench evaluation. If the error remains after scale P.C. board replacement or repair attempts, contact Hill-Rom Technical Support.

Clinical Use Tip

Never troubleshoot this issue on an active patient if the work could affect patient safety, bed position, power reliability, or scale accuracy. Move the patient to a backup bed or use another approved weighing method first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Hillrom VersaCare bed reported with Err1 on scale display. Scale function unavailable during attempted use."

Cause

What was observed during troubleshooting.

Example:
"Verified Err1 on scale display. Checked AC power, cycled bed power with 10-second wait, pressed siderail Enable key, and inspected for external scale interference. Error cleared after reset, indicating temporary scale electronics fault."

Resolution

What action was taken.

Example:
"Cycled bed power, enabled siderail controls, confirmed Err1 cleared, performed bed function checks, verified scale display operation, and returned bed to service."

Helpful Details to Include (If Known)

Final Thought

Err1 should be handled as a scale-system fault until proven otherwise. Start with patient safety, verify the displayed error, rule out power and external causes, reset the bed properly, and escalate when the error remains. Strong documentation protects patient care and repair history.

That is successful troubleshooting.