Hillrom VersaCare

Error: Err2 Scale Display Error

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

VersaCare

What This Guide Helps With

This guide helps troubleshoot Err2 on the Hillrom VersaCare scale display when the bed scale cannot function normally, may require recalibration, or cannot be trusted for accurate patient weight measurement.

Err2 affects the scale system and may prevent reliable patient weight documentation. Normal bed movement may still work, but the scale should not be used for clinical decisions until the issue is corrected and verified.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Cycle Bed Power

Check Power and Basic External Conditions

Attempt Scale Calibration if Err2 Remains

Inspect for Visible Frame or Scale System Problems

Correct Any Visible Problems Found

Determine Whether the Issue Is Persistent

Escalate for Repair Evaluation

If the Problem Persists

If Err2 remains after power cycling, pressing Enable, calibration, external power checks, and visible frame or cable inspection, common external causes have been ruled out.

The bed should be removed from service, labeled Out of Service, and sent for repair or bench evaluation. Escalate for scale P.C. board replacement or Hill-Rom Technical Support if needed.

A bed with an unresolved scale error should not be returned to service for patient weighing.

Clinical Use Tip

Never troubleshoot this issue on an active patient when frame movement, cable inspection, or scale repair is needed. Move the patient to a backup bed or use another approved weighing method before continuing troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Hillrom VersaCare bed reported Err2 on the scale display. Scale function could not be verified for patient weight use."

Cause

What was observed during troubleshooting.

Example:
"Err2 remained after initial scale display verification. Power cycle, Enable key check, calibration attempt, frame inspection, cable inspection, and weigh frame interference check were performed. Persistent error indicates likely scale system fault."

Resolution

What action was taken.

Example:
"Removed bed from service. Labeled Out of Service. Performed power cycle and scale troubleshooting. No correctable external issue found. Escalated for scale P.C. board replacement and further repair evaluation."

Helpful Details to Include (If Known)

Final Thought

Err2 on a Hillrom VersaCare bed should be handled as a scale system fault until proven otherwise. Start with patient safety, verify the exact display error, rule out power and setup issues, attempt calibration, inspect for visible mechanical or cable problems, and escalate when the error remains. Clear documentation protects the patient, the technician, and the equipment history.

That is successful troubleshooting.