Hillrom VersaCare

Error: Err5 Scale Display Error

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

VersaCare

What This Guide Helps With

This guide helps troubleshoot Err5 on the Hillrom VersaCare bed scale display. Err5 may prevent normal scale operation and can make the bed scale unreliable for patient weight measurement. The condition may be related to European scale configuration settings or scale P.C. board failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check External and Simple Causes First

Check for Obvious Physical or Environmental Issues

Confirm the Error Context

Check DIP Switch Configuration

If the Error Clears, Perform Function Checks

If Err5 Does Not Clear

If the Problem Remains After Scale Board Replacement

If the Problem Persists

If Err5 remains after verifying stable power, checking for obvious damage, confirming DIP switch #4 and DIP switch #2 are OFF, and replacing the scale P.C. board if authorized, external and common causes have been ruled out.

Remove the bed from service, label it Out of Service, document the failure clearly, and send the bed for repair or bench evaluation. Escalate to Hill-Rom Technical Support or qualified repair support if the error remains after the specified correction path.

Clinical Use Tip

Never troubleshoot on an active patient when service access, configuration changes, or internal scale components are involved. Move the patient to a backup bed or use another approved weighing method first if accurate patient weight is needed.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Hillrom VersaCare bed scale display showing Err5. Scale function unavailable or not reliable for patient weight measurement."

Cause

What was observed during troubleshooting.

Example:
"Err5 confirmed on scale display. Power and visible condition checked. Manual identifies Scale Error 4 and Scale Error 5 as European Scale Only conditions. DIP switch #4 and DIP switch #2 required OFF position verification."

Resolution

What action was taken.

Example:
"Set DIP switch #4 and DIP switch #2 to OFF. Err5 cleared. Performed bed and scale function checks. Bed passed checks and was returned to service."

Helpful Details to Include (If Known)

Final Thought

Err5 on a Hillrom VersaCare bed should be approached as a scale configuration or scale electronics issue. Start with patient safety, confirm the displayed error, rule out simple power and physical causes, verify the DIP switch settings, and only then escalate to scale board replacement or Hill-Rom support. Clear documentation protects the patient, the technician, and the facility.

That is successful troubleshooting.