Hologic Dimensions Series

Compression paddle sensor or force calibration errors

On this page

Asset Type

Mammography System

Manufacturer

Hologic

Model

Dimensions Series

What This Guide Helps With

Identifies external causes of paddle sensor or force calibration errors, including misalignment, incorrect paddles, obstructions, or system readiness issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected Outcome: Safe environment with no risk of unintended compression.

Observe Error Message and Paddle Behavior

Expected Outcome:

Why: Helps determine if the issue is mechanical, sensor-related, or system logic.

Verify Compression Paddle Installation

Expected Outcome: System recognizes paddle and clears sensor error.

Why: Improper seating or incorrect paddle type is a common cause of sensor faults.

Inspect Paddle and Mounting Interface

Expected Outcome: Clear interface allows proper sensor detection.

Why: Physical interference can prevent accurate paddle recognition or force sensing.

Check for Obstructions in Compression Path

Expected Outcome: Unobstructed motion allows normal force calibration.

Why: Unexpected resistance can trigger force calibration errors.

Verify System Ready and Calibration State

Expected Outcome: System completes calibration or clears fault.

Why: Some errors occur if calibration was interrupted or not completed properly.

Power Cycle the System

Expected Outcome: Temporary faults or sensor communication issues are cleared.

Why: Resets system logic and reinitializes paddle detection.

Test with Alternate Paddle (If Available)

Expected Outcome:

Why: Helps isolate whether the problem is accessory-specific or system-related.

If the Problem Persists

If all external causes have been ruled out, the issue is likely internal (sensor, actuator, or calibration system failure).

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Clinical Use Tip

Do not troubleshoot compression or force-related issues with a patient in the unit. Always transfer the patient before testing paddle movement or calibration.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Compression paddle error displayed and system would not allow proper compression."

Cause

What was observed during troubleshooting.

Example:
"Compression paddle not fully seated, preventing proper sensor recognition."

Resolution

What action was taken.

Example:
"Reseated paddle and verified normal operation with successful compression test."

Helpful Details to Include (If Known)

Final Thought

Start with the paddle itself—most issues are related to installation, alignment, or obstruction. Maintain a logical approach and prioritize patient safety. Escalate when internal calibration or sensor issues are suspected, and document clearly to support efficient repair.

That is successful troubleshooting.