Hologic Dimensions Series

Software Reconstruction or DICOM / PACS Transfer Failures

On this page

Asset Type

Mammography System

Manufacturer

Hologic

Model

Dimensions Series

What This Guide Helps With

Identifies external causes of reconstruction delays or PACS transfer failures including network issues, storage limits, workflow interruptions, or system communication problems.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected Outcome:
No risk of data loss or patient care disruption.

Check Reconstruction Status and Queue

Expected Outcome:
Images complete reconstruction or display a clear error.

Why:
Backlogged or stalled reconstruction prevents images from being sent to PACS.

Verify Available System Storage

Expected Outcome:
At least 10–15% free disk space available.

Why:
Insufficient storage can stop both reconstruction and DICOM export processes.

Confirm Network Connectivity

Expected Outcome:
Stable network connection is present.

Why:
DICOM transfer requires active network communication.

Inspect Network Cabling and Ports

Expected Outcome:
Secure connections restore communication if cabling was loose.

Why:
Loose or damaged cables commonly cause intermittent transfer failures.

Verify DICOM Destination Configuration

Expected Outcome:
Correct configuration allows successful communication with PACS.

Why:
Incorrect DICOM settings prevent image routing even if the network is active.

Check for PACS or Server Availability Issues

Expected Outcome:
If PACS is down, issue is external to the modality.

Why:
System may be functioning correctly but unable to send to an unavailable destination.

Restart DICOM/Acquisition Services or Workstation

Expected Outcome:
System clears stalled processes and resumes reconstruction/transfer.

Why:
Software processes can hang and require a restart to recover.

Attempt Manual Resend of Studies

Expected Outcome:
Images successfully transmit to PACS.

Why:
Confirms whether the issue was temporary or persistent.

If the Problem Persists

If reconstruction failures or DICOM transfer issues continue after verifying storage, network, configuration, and PACS availability, the issue is likely internal (software corruption, database issues, or system communication faults).

Remove the system from service, label it Out of Service, and escalate for vendor service or advanced bench evaluation.

Knowing when to stop prevents data loss and workflow disruption.

Clinical Use Tip

Do not troubleshoot during an active patient exam. Ensure all images are safely stored or backed up before performing restarts or system checks.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Images not reconstructing and failing to send to PACS from Hologic Dimensions system."

Cause

What was observed during troubleshooting.

Example:
"Network cable partially disconnected causing intermittent communication failure."

Resolution

What action was taken.

Example:
"Reseated Ethernet connection and verified successful reconstruction and PACS transfer."

Helpful Details to Include (If Known)

Final Thought

Effective troubleshooting starts with protecting patient data and workflow, then systematically verifying storage, network, and configuration before escalation. Proper documentation ensures repeat issues are resolved faster and more accurately.

That is successful troubleshooting.