Hologic Dimensions Series

X-ray generator or exposure synchronization faults

On this page

Asset Type

Mammography System

Manufacturer

Hologic

Model

Dimensions Series

What This Guide Helps With

Identifies external causes of exposure failures or sync faults, including interlocks, readiness conditions, positioning, and system communication preventing X-ray generation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected Outcome:
Safe environment with no radiation risk.

Verify System “Ready” Status

Expected Outcome:
System indicates ready for exposure.

Why:
The generator will not fire unless all system conditions are satisfied.

Check Emergency Stop (E-Stop) Buttons

Expected Outcome:
Releasing an engaged E-stop restores exposure capability.

Why:
E-stops interrupt generator operation and synchronization.

Verify Room and Door Interlocks

Expected Outcome:
Closed door clears interlock-related exposure faults.

Why:
Open or faulty interlocks prevent X-ray generation for safety compliance.

Inspect Compression and Positioning Status

Expected Outcome:
Proper positioning allows system to enter exposure-ready state.

Why:
Improper positioning or compression prevents exposure synchronization.

Check Footswitch / Exposure Switch Operation

Expected Outcome:
Functional switch triggers exposure sequence.

Why:
A failed or disconnected switch can mimic generator faults.

Review System Error Messages and Logs

Expected Outcome:
Specific fault codes help determine if issue is interlock, communication, or generator-related.

Why:
Error codes guide whether the issue is external or requires escalation.

Perform a Controlled System Restart

Expected Outcome:
System reinitializes and clears temporary synchronization faults.

Why:
Startup timing or communication issues can affect generator readiness.

Verify Power Stability to the System

Expected Outcome:
Stable power supports normal generator operation.

Why:
Voltage instability can prevent generator initialization or firing.

Confirm Generator Warm-Up / Calibration Status (if applicable)

Expected Outcome:
Completing warm-up allows exposure to proceed normally.

Why:
Some systems inhibit exposure until generator readiness checks are complete.

If the Problem Persists

If all external causes have been ruled out, the issue is likely internal (generator, high-voltage components, or synchronization circuitry).

Remove the device from service, label it Out of Service, and send for vendor repair or qualified bench evaluation.

Knowing when to stop prevents unnecessary risk and ensures proper escalation.

Clinical Use Tip

Never troubleshoot exposure issues with a patient in position. Always move the patient to another system before testing or cycling exposures.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"System would not initiate exposure; generator fault displayed during exam."

Cause

What was observed during troubleshooting.

Example:
"Exposure footswitch found disconnected, preventing trigger signal to generator."

Resolution

What action was taken.

Example:
"Reconnected footswitch and verified normal exposure operation."

Helpful Details to Include (If Known)

Final Thought

Exposure faults should always be approached from a safety-first and logic-driven perspective. Rule out interlocks, accessories, and system readiness before assuming generator failure. Proper escalation and clear documentation protect both the patient and the equipment lifecycle.

That is successful troubleshooting.