ICU Medical CADD-Solis

Occlusion Alarms – Especially False Occlusions

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide provides a structured Clinical Engineering approach when an ICU Medical CADD-Solis infusion pump triggers occlusion alarms, including:

The focus is on logical, external, and easily verifiable checks before assuming tubing failure, pressure plate damage, or internal pressure sensor malfunction.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Inspect the Entire Tubing Path

3. Verify Proper Cassette Seating

4. Inspect the Pressure Plate and Channel

5. Confirm Correct and Compatible Tubing

6. Check Pump Orientation and Placement

7. Power Cycle the Pump

8. Test with Known-Good Components

9. Observe Alarm Pattern

If Occlusion Alarms Persist

After:

Persistent alarms suggest likely internal pressure sensing or detection issues.

At this point:

Knowing when to stop prevents unnecessary guesswork and protects patient care.

Clinical Use Tip

Do not perform occlusion troubleshooting on a pump delivering high-risk medications to an active patient.

Whenever possible:

Maintaining therapy continuity is more important than resolving the device issue in real time.

Work Order Documentation (CCR Method)

Complaint

What was reported by the clinical staff.

Example:

“Pump repeatedly alarming ‘Occlusion’ during infusion; tubing appears patent.”

Cause

What was observed during troubleshooting.

Example:

“Tubing inspected, no kinks noted. Cassette reseated. Pump power-cycled. Tested with known-good tubing and cassette. Occlusion alarm persisted. Suspect internal pressure sensor or pressure plate issue.”

OR

“Tubing found kinked under bed rail causing intermittent occlusion.”

Resolution

What action was taken.

Example:

Helpful Details to Include (If Known)

Final Thought

Most occlusion alarms are caused by external resistance — kinks, tubing tension, cassette misalignment, or debris in the pressure channel. True internal sensor failures are less common but should be suspected after logical external causes are ruled out.

A structured approach protects patient safety, preserves therapy continuity, prevents unnecessary escalation, and ensures accurate documentation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide