ICU Medical CADD-Solis

Air-in-Line Alarms

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide provides a structured Clinical Engineering approach when an ICU Medical CADD-Solis infusion pump triggers:

The focus is on logical, external, and easily verifiable checks before assuming cassette failure, tubing defects, or internal sensor malfunction.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Patient safety always takes priority over device troubleshooting.

2. Inspect Tubing and Cassette for Air

Even small bubbles near the sensor can trigger alarms.

If air is present:

3. Check All Tubing Connections

Loose connections can introduce air during infusion, especially at higher flow rates.

4. Confirm Proper Cassette Installation

A partially seated cassette can cause false air detection.

5. Verify Pump Orientation and Placement

Improper orientation or tension can affect sensor accuracy.

6. Power Cycle the Pump

Temporary sensor or logic glitches may clear after a reboot.

7. Test with Known-Good Tubing and Cassette

If alarms stop:

If alarms continue:

8. Observe and Document Alarm Pattern

Note the following:

Pattern recognition helps distinguish external setup issues from internal failures.

If Air-in-Line Alarms Persist

If alarms continue after ruling out:

The likely causes include:

These are not correctable through bedside troubleshooting.

Next steps:

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never attempt repeated resets or bypass alarms during active infusion without ensuring patient therapy is maintained.

If therapy is critical:

Air-in-line alarms are safety features and should not be overridden without verification.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:

“Pump repeatedly alarming for air-in-line; no visible air in tubing.”

Cause

What was observed during troubleshooting.

Example:

“Tubing inspected and re-primed; cassette reseated; alarm persists.”
“Replaced tubing and cassette; alarms continue. Suspected internal air sensor issue.”

Resolution

What action was taken.

Example:

“Removed from service, labeled Out of Service, sent for bench evaluation.”
“Replaced defective tubing and cassette; infusion resumed without alarms.”
“Power cycled and reseated cassette; issue resolved.”

Helpful Details to Include (If Known)

Final Thought

Most air-in-line alarms are caused by trapped air, loose connections, or cassette alignment issues. Internal sensor failures are less common but possible.

A structured, step-by-step approach:

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide