ICU Medical CADD-Solis

Alarm: Air in line detected

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the Air in line detected alarm on the ICU Medical CADD-Solis infusion pump. This alarm indicates that the pump detected air in the fluid path and may interrupt medication delivery until the condition is corrected.

The technical manual lists this as a high-priority alarm. Air detector sensitivity may vary by configuration. Low sensitivity detects a single bubble greater than 400 μL, high sensitivity detects a single bubble greater than 150 μL, and accumulated air is greater than 1 mL over 15 minutes nominal.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm

Check the Reservoir, Cassette, and Tubing for Air

Confirm the Cassette and Administration Set Are Installed Correctly

Inspect the Air Detector Sensor Area

Retest With a Properly Primed Set

If the Problem Persists

If the alarm continues after visible air has been removed, the cassette and tubing have been checked, the disposable set has been primed or replaced, and the air detector sensor slot is clean, remove the pump from service.

Label the pump Out of Service and send it for repair or bench evaluation. Persistent air-in-line alarms may require further inspection of the air detector system, pump configuration, or internal components by qualified service personnel.

Clinical Use Tip

Never troubleshoot an infusion pump on an active patient when medication delivery may be interrupted or unsafe. Move the patient to a backup pump or approved alternate delivery method first, then troubleshoot the removed pump.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ICU Medical CADD-Solis pump alarming Air in line detected during infusion."

Cause

What was observed during troubleshooting.

Example:
"Found air visible in administration set near cassette. Air detector sensor slot inspected and found clean with no debris."

Resolution

What action was taken.

Example:
"Pump removed from patient use. Disposable set was re-primed/replaced per facility procedure. Pump retested with properly primed set and alarm did not return."

Helpful Details to Include (If Known)

Final Thought

An air-in-line alarm should be treated as a patient safety issue first and a device troubleshooting issue second. Rule out visible air, priming problems, cassette seating, tubing setup, and sensor slot contamination before escalating the pump for repair.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide