ICU Medical CADD-Solis

Alarm: Battery depleted while delivering

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the Battery depleted while delivering alarm on an ICU Medical CADD-Solis infusion pump. This is a high-priority power alarm that occurs when battery power is too low to operate the pump. Delivery stops, a continuous variable-tone alarm sounds, and the pump requires restored power before it can safely continue operation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the Alarm Message

Restore Pump Power

Check External Power Accessories

Verify Delivery Status Before Restarting

Observe for Repeat Alarm Behavior

If the Problem Persists

If fresh AA batteries, a charged rechargeable battery pack, and AC power have been checked and the pump still reports depleted battery power or will not operate normally, external power causes have been ruled out. Remove the device from service, label it Out of Service, and send it for repair or bench evaluation. Do not return the pump to clinical use until power operation and delivery behavior are verified.

Clinical Use Tip

Never troubleshoot this alarm on an active patient if medication delivery may be interrupted or unsafe. Move the patient to a backup pump or approved alternate delivery method first, then troubleshoot the CADD-Solis pump after patient therapy is protected.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported ICU Medical CADD-Solis infusion pump displayed Battery depleted while delivering alarm and stopped delivery during use."

Cause

What was observed during troubleshooting.

Example:
"Confirmed high-priority battery depleted alarm. Pump battery power was too low to operate. Battery set was depleted and delivery had stopped."

Resolution

What action was taken.

Example:
"Replaced batteries and connected pump to AC power. Verified pump powered on normally, alarm cleared, and clinical staff confirmed delivery status before returning pump to use."

Helpful Details to Include (If Known)

Final Thought

A depleted battery alarm during delivery is a patient safety issue because the pump stops delivering therapy. Protect the patient first, restore reliable power, verify delivery status with clinical staff, and escalate the pump if normal power operation cannot be confirmed.

That is successful troubleshooting.