What This Guide Helps With

This guide helps troubleshoot the ICU Medical CADD-Solis Battery removed while delivering alarm. This is a high-priority alarm that indicates the pump detected battery removal while medication delivery was active. The pump may also generate a battery fallout alarm when the pump has been powered up long enough and is in run mode.

Because this alarm occurs during delivery, the main concern is confirming patient safety, restoring power, checking the battery compartment, and verifying the pump’s delivery status before returning it to use.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Restore Power to the Pump

Inspect the Battery Door

Check Battery Fit and Condition

Inspect External Battery Contacts

Verify Delivery Status Before Reuse

If the Problem Persists

If the batteries, battery door, battery fit, AC power source, and visible battery contacts have been checked and the alarm continues or returns, external causes have been ruled out.

Remove the pump from service, label it Out of Service, and send it for repair or bench evaluation. Do not return the pump to patient use if the battery door is damaged, the batteries do not stay seated, the contacts appear damaged, or the pump cannot reliably maintain power during delivery.

Clinical Use Tip

Never troubleshoot this alarm on an active patient if therapy may be interrupted. Move the patient to a backup pump or approved alternate delivery method first, then evaluate the CADD-Solis pump after patient care is protected.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ICU Medical CADD-Solis pump alarmed Battery removed while delivering during active infusion."

Cause

What was observed during troubleshooting.

Example:
"Batteries were loose in compartment and battery door was not fully latched. AC power restored pump operation and alarm cleared."

Resolution

What action was taken.

Example:
"Reinstalled batteries, verified battery door latch, inspected battery contacts, connected AC power, confirmed alarm cleared, and advised staff to verify delivery status before reuse."

Helpful Details to Include (If Known)

Final Thought

A battery removal alarm during delivery should always be treated as a patient safety event first and a power troubleshooting issue second. Confirm therapy safety, restore reliable power, inspect the battery compartment, and escalate the pump if the battery system cannot be trusted.

That is successful troubleshooting.