ICU Medical CADD-Solis

Alarm: Battery unusable while delivering

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the Battery unusable while delivering alarm on the ICU Medical CADD-Solis infusion pump. This is a high-priority alarm that may occur when the pump detects batteries that cannot provide proper power while medication delivery is active.

This condition can interrupt therapy, stop reliable pump operation, or prevent the pump from safely continuing delivery until the power source issue is corrected.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the Alarm

Use Only Approved Batteries

Check for Battery Damage or Poor Contact

Check the Battery Door and External Power Option

Verify Operation Before Returning to Use

If the Problem Persists

If approved batteries or the approved rechargeable battery pack have been installed, battery orientation has been verified, the battery door and contacts have been inspected, and AC power has been checked but the alarm continues, remove the pump from service.

Label the device Out of Service and send it for repair or bench evaluation. Possible external causes have been ruled out, and the pump may require further testing of the battery contacts, power path, battery door fit, or internal power circuitry by qualified service personnel.

Clinical Use Tip

Never troubleshoot this alarm on an active patient if therapy could be interrupted. Move the patient to a backup pump first, then inspect batteries, contacts, and power accessories away from active patient use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ICU Medical CADD-Solis infusion pump displayed Battery unusable while delivering alarm during use."

Cause

What was observed during troubleshooting.

Example:
"Pump had non-approved batteries installed. Battery contacts and battery compartment were inspected with no leakage or damage found."

Resolution

What action was taken.

Example:
"Installed approved AA alkaline batteries, verified battery orientation, confirmed alarm cleared, checked pump power behavior, and verified device ready for use per facility procedure."

Helpful Details to Include (If Known)

Final Thought

A battery unusable alarm during delivery should be treated as a patient safety and therapy continuity issue, not just a battery nuisance. Start with the approved power source, inspect for obvious battery or contact problems, verify stable operation, and escalate the pump if the alarm returns after external causes are ruled out.

That is successful troubleshooting.