ICU Medical CADD-Solis

Alarm: Disposable attached improperly

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the ICU Medical CADD-Solis Disposable attached improperly alarm. This is a high-priority alarm that occurs when the pump detects that the cassette or administration set is not correctly attached, latched, or locked to the pump.

The latch/lock sensor circuit helps prevent delivery when the disposable is not properly secured. If the disposable is attached incorrectly, medication delivery may stop or may be prevented from starting until the cassette or administration set is correctly installed.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Alarm Message

Remove and Reattach the Disposable

Inspect the Latch and Lock Area

Try a Known-Good Disposable Set if Available

Verify Pump Status Before Returning to Use

If the Problem Persists

If the alarm continues after the cassette or administration set has been removed and correctly reattached, a known-good compatible disposable has been tried, and the latch/lock area has been inspected for obvious external issues, remove the pump from service.

Label the device Out of Service and send it for repair or bench evaluation. Continued alarms may indicate a latch, lock, cassette detection, sensor, housing, or internal circuit issue that requires qualified service evaluation.

Clinical Use Tip

Never troubleshoot this alarm on an active patient while medication delivery is still needed. Move the patient to a backup pump or approved alternate delivery method first, then troubleshoot the CADD-Solis pump after patient therapy is protected.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ICU Medical CADD-Solis pump alarming Disposable attached improperly and would not allow delivery to continue."

Cause

What was observed during troubleshooting.

Example:
"Cassette was not fully seated and latch was not completely locked. Latch area inspected with no visible damage or debris found."

Resolution

What action was taken.

Example:
"Removed and reattached cassette correctly, verified cassette fully latched and locked, cleared alarm, and confirmed pump returned to normal operation before clinical use."

Helpful Details to Include (If Known)

Final Thought

The Disposable attached improperly alarm should be handled as a safety stop, not just a nuisance alarm. Start with patient safety, confirm the disposable is correctly installed, inspect the latch and cassette area, try a known-good disposable when appropriate, and escalate the pump if the alarm does not clear reliably.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide