ICU Medical CADD-Solis

Alarm: Disposable damaged

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the Disposable damaged alarm on an ICU Medical CADD-Solis infusion pump. The manual lists this as a high-priority alarm involving the attached disposable cassette or administration set. This alarm can prevent safe infusion delivery and should be treated as a cassette, tubing, latch, or set integrity concern until confirmed otherwise.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm

Inspect the Disposable Cassette and Administration Set

Check Cassette Seating and Latch Area

Replace Damaged or Questionable Disposable Components

Restart and Verify Safe Setup

If the Problem Persists

If the Disposable damaged alarm continues after the cassette, tubing, latch area, and administration set have been inspected and the disposable components have been replaced, external causes have been ruled out. Remove the device from service, label it Out of Service, and send it for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot this alarm on an active patient while infusion therapy is expected to continue. Move the patient to a backup pump or alternate approved infusion method first, then inspect the cassette, tubing, latch area, and administration set.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ICU Medical CADD-Solis infusion pump displayed Disposable damaged high-priority alarm during setup."

Cause

What was observed during troubleshooting.

Example:
"Inspected cassette, tubing, latch area, and administration set. Disposable cassette appeared damaged and alarm repeated with original set."

Resolution

What action was taken.

Example:
"Replaced disposable administration set, reprimed tubing, verified cassette seated properly, and confirmed alarm cleared before returning pump for clinical setup verification."

Helpful Details to Include (If Known)

Final Thought

A Disposable damaged alarm should be handled as a patient safety and disposable integrity issue first. Confirm the alarm, inspect the cassette and administration set, replace anything damaged or questionable, reprime the set, and only return the pump to use after safe setup is verified. If the alarm persists after external causes are ruled out, remove the pump from service and document the issue clearly.

That is successful troubleshooting.