ICU Medical CADD-Solis

Alarm: Disposable detached while delivering

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the ICU Medical CADD-Solis high-priority alarm Disposable detached while delivering. This alarm indicates the pump detected that the disposable cassette or set became detached, unlatched, or improperly secured while therapy was being delivered.

The latch/lock sensor circuit allows the pump microprocessor to stop delivery and alarm if the set is unlatched during delivery. Because medication or fluid delivery may be interrupted, this alarm should be treated as a patient care concern until the setup is corrected and therapy status is verified.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm

Inspect the Disposable Cassette or Set Attachment

Check the Latch and Lock Status

Reconnect or Replace the Disposable

Confirm Operation Before Returning to Use

If the Problem Persists

If the alarm continues after the disposable has been reattached or replaced, external causes such as incorrect cassette seating, disposable damage, latch position, tubing tension, and lock status have been ruled out.

Remove the device from service, label it Out of Service, and send it for repair or bench evaluation. A recurring alarm may indicate a latch, lock, sensor, or pump detection circuit issue that should not be ignored.

Clinical Use Tip

Never troubleshoot on an active patient if doing so could interrupt therapy, disturb the infusion setup, or create medication delivery risk. Move the patient to a backup pump or approved alternate delivery method first, then evaluate the CADD-Solis pump when it is safe to do so.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ICU Medical CADD-Solis pump displayed high-priority alarm Disposable detached while delivering during active infusion."

Cause

What was observed during troubleshooting.

Example:
"Disposable cassette was not fully seated and latch status was not secure. Existing disposable was inspected and replaced."

Resolution

What action was taken.

Example:
"Replaced disposable set, confirmed cassette fully attached, verified latch and lock status, alarm cleared, and pump was returned to clinical staff for therapy verification."

Helpful Details to Include (If Known)

Final Thought

A disposable detached alarm during delivery should be handled as both a pump issue and a therapy interruption concern. Start with patient safety, confirm the cassette and latch externally, replace the disposable when needed, and escalate the pump if the alarm returns after proper setup.

That is successful troubleshooting.