ICU Medical CADD-Solis

Alarm: Disposable locked but not latched

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the ICU Medical CADD-Solis high-priority alarm Disposable locked but not latched. This alarm indicates the pump is detecting a cassette lock status and latch status that do not match a valid safe delivery condition.

The disposable cassette may be locked without being properly latched, seated incorrectly, or not fully secured to the pump. Because this condition can affect safe fluid delivery, the pump should not be used for active infusion until the disposable is correctly attached and the alarm is cleared.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the Reported Alarm

Inspect the Disposable Cassette Position

Latch the Disposable Before Locking

Check for Disposable or Cassette Damage

Verify Alarm Clears After Reseating

If the Problem Persists

If the alarm remains after the disposable is unlocked, properly latched, seated correctly, and locked, external setup issues have been ruled out.

Remove the infusion pump from service, label it Out of Service, and send it for repair or bench evaluation. Possible causes may include a worn latch mechanism, cassette lock detection issue, damaged housing, or internal sensor/switch fault.

Clinical Use Tip

Never troubleshoot this alarm on an active patient while the infusion is expected to continue. Move the patient therapy to a backup pump first, then inspect the disposable, cassette latch, and lock condition once it is safe to do so.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ICU Medical CADD-Solis pump displayed high-priority alarm: Disposable locked but not latched."

Cause

What was observed during troubleshooting.

Example:
"Disposable cassette was locked before being fully latched and seated to the pump. Cassette latch and lock status did not match safe delivery condition."

Resolution

What action was taken.

Example:
"Unlocked cassette lock, reseated disposable cassette, properly latched disposable, then locked cassette. Alarm cleared and pump completed setup normally."

Helpful Details to Include (If Known)

Final Thought

A disposable latch or lock alarm should be handled as a safety-related setup issue, not just a nuisance alarm. Confirm the cassette is properly seated, latched, and locked before returning the pump to use. If the pump continues to detect an invalid cassette condition after external checks, remove it from service and document the complaint, cause, and resolution clearly.

That is successful troubleshooting.