ICU Medical CADD-Solis

Alarm: Disposable type high flow administration set not allowed

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the CADD-Solis alarm Disposable type high flow administration set not allowed. The manual lists this as a high-priority alarm. It can occur when a high-volume administration set is attached to a pump model or therapy setup that does not allow that disposable type.

High-volume administration sets are for model 2110 only. If a high-volume administration set is attached to a model 2100, the pump will alarm and will not run. This prevents the pump from delivering therapy with an incorrect disposable set configuration.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Alarm Message

Check the Administration Set Type

Confirm the Pump Model and Therapy Setup

Replace the Disposable Set if Needed

Confirm Alarm Clears and Pump Runs Normally

If the Problem Persists

If the correct disposable set is installed, the pump model and therapy setup have been verified, and the alarm still returns, external setup causes have been ruled out.

Remove the pump from service. Label it Out of Service. Send the pump for repair, inspection, or bench evaluation according to facility procedure.

Clinical Use Tip

Never troubleshoot an infusion pump on an active patient when therapy delivery may be interrupted or uncertain. Move the patient to a backup pump or approved alternate delivery method first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"CADD-Solis pump displayed high-priority alarm: Disposable type high flow administration set not allowed. Pump would not run."

Cause

What was observed during troubleshooting.

Example:
"High-volume administration set was attached to a pump setup that did not allow that disposable type. Pump model and disposable set compatibility were reviewed."

Resolution

What action was taken.

Example:
"Incorrect disposable set was removed and correct administration set was installed for the pump model and programmed therapy. Alarm cleared and pump operation was verified with clinical staff."

Helpful Details to Include (If Known)

Final Thought

This alarm is usually a setup and compatibility issue, not something to work around. Confirm the pump model, confirm the disposable set, protect the patient, and escalate the pump only after the correct external setup has been verified.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide