ICU Medical CADD-Solis

Alarm: Disposable type high flow administration set required

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the ICU Medical CADD-Solis alarm Disposable type high flow administration set required. This is a high-priority alarm related to the programmed delivery rate, pump model, and administration set type.

Programmed delivery rates above 250 mL/hr require a high-volume administration set and a CADD-Solis 2110 pump. If the required setup is not present, the pump may alarm and prevent therapy from continuing as programmed.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the Alarm Message

Verify Pump Model

Verify the Administration Set

Check the Programmed Therapy Setup

Restart and Recheck

If the Problem Persists

If the correct high-volume administration set is installed, the correct CADD-Solis 2110 pump is being used, and the therapy setup matches the prescription but the alarm continues, external setup causes have been ruled out.

Remove the pump from service, label it Out of Service, and send it for repair or bench evaluation. Do not return the pump to clinical use until the alarm behavior has been evaluated and corrected.

Clinical Use Tip

Never troubleshoot an infusion pump setup on an active patient when therapy delivery may be interrupted or changed. Move the patient to a backup pump or approved alternate delivery method first, then troubleshoot the removed device and setup.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"CADD-Solis pump displayed high-priority alarm Disposable type high flow administration set required during therapy setup."

Cause

What was observed during troubleshooting.

Example:
"Programmed delivery rate was above 250 mL/hr and the installed disposable set was not the required high-volume administration set."

Resolution

What action was taken.

Example:
"Clinical staff replaced setup with required high-volume administration set and verified use with CADD-Solis 2110 pump. Alarm cleared and pump function was verified before return to use."

Helpful Details to Include (If Known)

Final Thought

This alarm should be handled as a therapy setup and compatibility issue first. Confirm patient safety, verify the pump model, confirm the administration set, check the ordered rate, and escalate the pump only after external setup problems have been ruled out.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide