ICU Medical CADD-Solis

Alarm: Disposable type invalid

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the Disposable type invalid alarm on the ICU Medical CADD-Solis infusion pump. The manual lists this as a high-priority alarm. The pump uses a cassette type sensor circuit with three pins that interface with molded features in the attached administration set to identify the disposable type. If the pump cannot recognize the cassette or set type correctly, infusion therapy may be interrupted or prevented from starting.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm

Confirm Disposable Compatibility

Replace the Disposable

Inspect External Cassette Seating and Pump Contact Area

Test With a Known-Good Compatible Disposable

Check for User Setup or Loading Issues

If the Problem Persists

If compatible disposables have been confirmed, the disposable has been replaced, cassette seating has been checked, and the alarm still appears with a known-good set, external causes have been ruled out. Remove the pump from service, label it Out of Service, and send it for repair or bench evaluation. The pump may require evaluation of the cassette type sensor pins, sensor circuit, latch area, or related internal components by qualified service personnel.

Clinical Use Tip

Never troubleshoot this alarm on an active patient. Move the patient to a backup pump or alternate approved infusion method first. A high-priority disposable recognition alarm can interrupt infusion therapy and should be treated as a patient safety issue until resolved.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ICU Medical CADD-Solis pump alarming Disposable type invalid after administration set was attached."

Cause

What was observed during troubleshooting.

Example:
"Confirmed high-priority Disposable type invalid alarm. Original cassette was not recognized by pump. Verified compatible set and inspected cassette seating area for visible obstruction."

Resolution

What action was taken.

Example:
"Replaced disposable with known-good compatible CADD-Solis administration set. Pump recognized cassette type and alarm cleared. Pump passed functional check and was returned to service."

Helpful Details to Include (If Known)

Final Thought

The Disposable type invalid alarm should be handled with a safety-first approach because it can prevent correct infusion setup or interrupt therapy. Start with the disposable, compatibility, cassette seating, and external contact area before suspecting pump failure. If the pump does not recognize multiple known-good compatible disposables, remove it from service and document the complaint, cause, and resolution clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide