ICU Medical CADD-Solis

Alarm: Downstream occlusion

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot a Downstream occlusion alarm on an ICU Medical CADD-Solis infusion pump. This alarm occurs when the pump detects downstream pressure or restriction in the patient line. The pump may stop medication delivery and issue visual and audible alarms until the occlusion is corrected.

The technical information notes that the downstream occlusion analog sensor trips when tubing inflation corresponds to 18 ± 9 psi. The listed maximum infusion pressure is 27 psi / 1.86 bar.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm

Check the Tubing and Clamps

Check for Patient-Line or Catheter Obstruction

Check Filters, Connectors, and Accessories

Check for High Back Pressure

Restart Only After Correction

If the Problem Persists

If closed clamps, kinked tubing, blocked catheter concerns, clogged filters, high back pressure, patient-line obstruction, and accessory issues have been ruled out, remove the infusion pump from service.

Label the pump Out of Service and send it for repair or bench evaluation. Do not return the pump to clinical use if the Downstream occlusion alarm repeats without a confirmed external cause.

Clinical Use Tip

Never troubleshoot this alarm on an active patient in a way that could delay therapy, disconnect medication delivery unsafely, or expose the patient to pressure release risk. Move the patient to a backup pump or alternate approved delivery method first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ICU Medical CADD-Solis infusion pump alarming Downstream occlusion during infusion and stopping delivery."

Cause

What was observed during troubleshooting.

Example:
"Found downstream tubing kinked near patient connection, causing elevated line pressure and repeated occlusion alarm."

Resolution

What action was taken.

Example:
"Tubing restriction corrected by clinical staff, line pressure relieved safely, pump restarted and observed with no repeat Downstream occlusion alarm. Pump returned to service."

Helpful Details to Include (If Known)

Final Thought

A Downstream occlusion alarm should be handled as a medication delivery interruption first and an equipment problem second. Start with patient safety, check the external fluid path, confirm the line is open, and escalate the pump only after tubing, catheter, filter, accessory, and back-pressure causes have been ruled out.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide