ICU Medical CADD-Solis

Alarm: Key stuck

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot a Key stuck alarm on the ICU Medical CADD-Solis infusion pump. The manual lists this as a high-priority alarm. A stuck, pressed, damaged, or contaminated keypad button may prevent normal pump operation and should be treated as a patient safety concern.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Alarm

Inspect the Keypad Externally

Clean the Keypad Area

Retest Keypad Operation

If the Problem Persists

If the keypad has been checked for stuck, pressed, damaged, or contaminated keys and the alarm still returns, external causes have been ruled out as much as possible. Remove the pump from service, label it Out of Service, and send it for repair or bench evaluation. Do not return the pump to patient use unless the keypad operates normally and the high-priority alarm is resolved.

Clinical Use Tip

Never troubleshoot this alarm on an active patient while the pump is responsible for medication or fluid delivery. Move the patient infusion to a backup pump first, then inspect and clean the CADD-Solis pump when it is safe to do so.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ICU Medical CADD-Solis infusion pump reported Key stuck high-priority alarm and keypad would not operate normally."

Cause

What was observed during troubleshooting.

Example:
"Inspected keypad for stuck, pressed, damaged, or contaminated keys. Keypad showed abnormal key response after cleaning and retest."

Resolution

What action was taken.

Example:
"Removed pump from service, labeled Out of Service, and sent for repair or bench evaluation due to unresolved Key stuck alarm."

Helpful Details to Include (If Known)

Final Thought

A Key stuck alarm should not be ignored because the keypad is part of the pump control interface. Start with patient safety, verify the alarm, inspect and clean the keypad externally, and remove the pump from service if normal keypad operation cannot be confirmed. Clear documentation helps show what was checked and why the pump was escalated.

That is successful troubleshooting.