ICU Medical CADD-Solis

Alarm: Pressure sensor faulty

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot a Pressure sensor faulty alarm on the ICU Medical CADD-Solis infusion pump. The manual lists this as a high-priority alarm. The pump uses occlusion sensors in the pumping mechanism to detect pressure-related conditions, so a pressure sensor fault may affect the pump's ability to reliably monitor pressure or detect occlusion-related issues.

This alarm should be treated seriously because pressure monitoring is part of safe infusion delivery. If the alarm persists after basic external checks, the pump should be removed from service and sent for repair or bench evaluation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm

Check the Cassette and Tubing Setup

Inspect the Pumping Mechanism Area

Power Cycle and Recheck

Check for Obvious External Damage

If the Problem Persists

If the alarm persists after cassette setup, tubing, debris, contamination, and visible external damage have been checked, remove the pump from service. External causes have been ruled out, and the pump may have a pressure sensor, pumping mechanism, or related internal fault.

Label the pump Out of Service and send it for repair or bench evaluation. Service or repair must be performed by Smiths Medical, ICU Medical, or an authorized service agent according to facility policy.

Clinical Use Tip

Never troubleshoot this alarm on an active patient. A Pressure sensor faulty alarm may affect pressure or occlusion detection. Move the patient therapy to a backup pump first, then evaluate the affected pump only after it is safe to remove from clinical use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported ICU Medical CADD-Solis infusion pump displayed high-priority alarm: Pressure sensor faulty."

Cause

What was observed during troubleshooting.

Example:
"Verified Pressure sensor faulty alarm. Inspected cassette seating, tubing setup, pressure plate area, pumping mechanism area, and exterior housing. No user-correctable external cause found."

Resolution

What action was taken.

Example:
"Removed pump from service, labeled Out of Service, and sent for authorized repair/bench evaluation due to persistent pressure sensor fault."

Helpful Details to Include (If Known)

Final Thought

A high-priority pressure sensor alarm on an infusion pump is not something to work around. Confirm the setup, check for debris or visible damage, rule out external causes, then remove the pump from service if the alarm persists. Safe escalation and clear CCR documentation protect the patient, the clinical team, and the technician.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide