ICU Medical CADD-Solis

Alarm: Rechargeable battery end of life

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the high-priority Rechargeable battery end of life alarm on the ICU Medical CADD-Solis infusion pump. This alarm indicates the rechargeable lithium-ion battery pack has reached the end of its expected service life or can no longer be relied on for normal pump operation.

The rechargeable battery pack is rated for a minimum of 500 charge/discharge cycles. Once this alarm appears, the battery pack should not be trusted for continued clinical use until the issue is corrected.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm

Check External Power and Charging Setup

Inspect the Rechargeable Battery Pack

Replace the Rechargeable Battery Pack

Verify Pump Operation After Battery Replacement

If the Problem Persists

If the alarm continues after the rechargeable battery pack has been replaced and external power, charging setup, battery condition, and installation have been checked, external causes have been ruled out.

Remove the infusion pump from service, label it Out of Service, and send it for repair or bench evaluation. Do not return the pump to clinical use until the battery alarm and charging/battery operation have been verified as corrected.

Clinical Use Tip

Never troubleshoot this alarm on an active patient if it could interrupt therapy. Move the patient to a backup pump first, then troubleshoot the CADD-Solis pump after it is safely removed from patient use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported ICU Medical CADD-Solis pump displayed high-priority Rechargeable battery end of life alarm."

Cause

What was observed during troubleshooting.

Example:
"Confirmed rechargeable battery end of life alarm. Rechargeable lithium-ion battery pack reached end-of-life condition and required replacement. Battery pack inspected for damage and wear."

Resolution

What action was taken.

Example:
"Removed pump from use, replaced rechargeable battery pack, verified pump powered on normally, confirmed alarm cleared, and returned pump to service after operational check."

Helpful Details to Include (If Known)

Final Thought

A rechargeable battery end-of-life alarm should be treated as a battery reliability issue, not just a nuisance alarm. Protect the patient first, confirm the alarm, inspect the battery pack, replace it with the correct approved battery, verify the alarm clears, and document the work clearly using CCR.

That is successful troubleshooting.