ICU Medical CADD-Solis

Alarm: Remote dose cord key stuck

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the high-priority Remote dose cord key stuck alarm on an ICU Medical CADD-Solis infusion pump. The remote dose cord allows the patient to request a PCA or PCEA dose when attached, so a stuck or unreliable button can affect patient-controlled dosing and pump operation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm

Inspect the Remote Dose Cord Button

Check the Remote Dose Cord Connection

Test With a Known Good Remote Dose Cord

Check for Cleaning or Handling Damage

If the Problem Persists

If the remote dose cord has been inspected, the button is not physically stuck, the connector has been checked, and a known good remote dose cord does not clear the alarm, remove the pump from service. Label the device Out of Service and send it for repair or bench evaluation. External causes such as a stuck patient dose button, damaged remote dose cord, poor connection, and accessory failure should be ruled out before internal pump evaluation.

Clinical Use Tip

Never troubleshoot this alarm on an active patient without clinical approval. For PCA or PCEA therapy, move the patient to a backup pump or approved alternate therapy first, then evaluate the remote dose cord and pump outside active patient use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported ICU Medical CADD-Solis pump displayed high-priority alarm: Remote dose cord key stuck during PCA setup."

Cause

What was observed during troubleshooting.

Example:
"Remote dose cord button was sticking and did not return reliably after being pressed. Pump alarm cleared when tested with known good remote dose cord."

Resolution

What action was taken.

Example:
"Removed faulty remote dose cord from service, installed replacement remote dose cord, verified alarm cleared, and returned pump to service after functional check."

Helpful Details to Include (If Known)

Final Thought

A Remote dose cord key stuck alarm should be treated seriously because the remote dose cord is part of the patient-controlled dosing workflow. Start with patient safety, inspect the dose cord and button, replace questionable accessories, avoid twisting the connector during removal, and escalate the pump if the alarm does not clear with a known good cord.

That is successful troubleshooting.